WARNING
•
Access to power switch and power cord with plug. Do not
block access to the power switch and power cord. The power
switch must always be easy to access. The power cord with
plug must always be easy to disconnect.
•
Protective ground. The product must always be connected
to a grounded power outlet.
•
Supply voltage. Before connecting the power cord, make sure
that the supply voltage at the wall outlet corresponds to the
marking on the instrument.
•
High voltage. The mains cable must only be connected by
authorized service personnel. Faulty connection might result
in live system parts that can give lethal electric shocks.
•
Installing the computer. The computer must be installed and
used according to the instructions provided by the manufac-
turer of the computer.
CAUTION
•
The wheels of ÄKTA ready should be locked during normal use.
The wheels should be unlocked only when moving the unit.
•
Make sure that all tubing, hoses and cables are placed so that
the risk of tripping accidents is minimized.
•
The product is designed for indoor use only.
•
Do not use the product in a dusty atmosphere or close to
spraying water.
•
Make sure that correct air pressure is always maintained. Too
high or too low air pressure may be hazardous and may cause
erroneous results and leakage.
•
If the system is re-configured, the system label must be updat-
ed accordingly. Use a permanent ink pen, or the stickers sup-
plied with the system.
18
ÄKTA ready Operating Instructions 28960345 AD
2 Safety instructions