3.3
Flow chart
Chart and components
The liquid flow through ÄKTA ready is illustrated in the chart below.
(Black: standard components; Red: optional components)
XV-056
XV-055
XV-054
XV-053
XV-052
XV-051
XV-001
XV-002
XV-003
XV-004
XV-005
XV-006
PT-111
PT-112
PT-113
AE-121 TE-161 AE-131
FE-141 CE-101
AT-221
HV-301
P-201
A
B
C
D
P-202
The table below specifies the maximum operating pressure the flow path could be
pressurized to with the pressure upgrade strategy in the different parts of the system.
Pressure
Description
Part
0.6 bar
Low pressure, over the Inlet Manifold
A
5.0 bar
High pressure, over the air trap manifold and column
B
2.0 bar
Medium pressure
C
0.95 bar
Medium pressure, over the outlet manifold
D
The table below describes the system components shown in the flow chart.
Function
Part
Inlet 1 to 6, grouped in Inlet Manifold
XV-001 to XV-006
ÄKTA ready Operating Instructions 28960345 AD
45
3 System description
3.3 Flow chart