Positioning the system
The area required at the production location for ÄKTA ready is 280 x 300 cm,
allowing for 100 cm of free space on all sides of the system. The
ÄKTA ready gradient needs slightly more free space (280 x 316 cm).
Note:
The space is required to make room for containers and hoses, and to allow for easy ac-
cess to the power switch and to connectors on the back of the unit. The back door of
the cabinet must furthermore be accessible for service.
WARNING
Access to power switch and power cord. The power switch must
always be easy to access. The power cord must always be easy to
disconnect.
ÄKTA ready arrives preassembled in the transport crate.
No specific assembly prior to site installation is required.
After positioning the system at its designated location, lock
the wheels (figure).
NOTICE
Check that the Pressure Control Regulator for the system air supply
is set to the pressure stated in the specifications.