Function
Part
Outlet 1 to 6, grouped in Outlet Manifold
XV-051 to XV-056
3 Pressure sensors cells
PT-111 to PT-113
pH electrode (optional)
AE-121
Flow meter cell
FE-141
Temperature sensor
TE-161
Conductivity sensor
CE-101
UV cell
AE-131
Air trap
AT-221
Air vent valve
HV-301
System pump
P-201
Gradient pump (optional)
P-202
See
Section 3.4 Equipment, on page 48
for more information about the parts in the table
above.
The gradient pump (P202) is described in
Chapter 7 ÄKTA ready gradient, on
.
Note:
From inlet to column
Description
Stage
Inlet valves open the appropriate inlet (XV-001 to XV-006 in chart) for sample
or buffer.
1
The system pump (P-201) and the gradient pump (P-202) deliver the liquid
to the column via a pressure sensor cell (PT-111) and, if preferred, via the
air trap (AT-221), where air in the liquid is removed.
2
There are two sets of valves between the pump and the column:
3
•
the air trap manifold, which allows for bypassing the air trap, and
•
the column manifold, which allows for bypassing the column.
46
ÄKTA ready Operating Instructions 28960345 AD
3 System description
3.3 Flow chart