Action
Step
Select P202Gain under Pump, click Execute when flow signal is stable at set
flow rate.
5
Select FlowCompensation under Pump and click ON and let the calibration
procedure compensate for differences between the capacity of pumps A
(P-201) and B (P-202).
Note:
FlowCompensation is default set to OFF and has to be activated in each run.
6
106
ÄKTA ready Operating Instructions 28960345 AD
7 ÄKTA ready gradient
7.3 Calibrate and operate the ÄKTA ready gradient pump