![GE AKTA ready Скачать руководство пользователя страница 217](http://html.mh-extra.com/html/ge/akta-ready/akta-ready_operating-instructions-manual_128010217.webp)
Remark
Sign
Check
Step
UNICORN installed and
working (see UNICORN doc-
umentation)?
3 Installation,
computer
Bar code scanner installed
(see instructions in the
scanner´s Quick Start
Guide)?
ÄKTA ready strategy soft-
ware installed and verified
(see UNICORN documenta-
tion)?
Test fluids prepared.
4 Preparation,
Installation
test
Inlet/Outlet containers
available?
Flow kit for testing avail-
able.
Performed installation test
(details below)?
5 Installation
test
Installation test checklist
Calibration, pressure sensors
Perform pressure sensor calibration as described in
Calibrating pressure sensors, on
.
Remark
Sign
Perform
Pressure
sensor
Calibration at 0 bar g
PT-111
Calibration at 5 bar g
Calibration at 0 bar g
PT-112
Calibration at 5 bar g
Calibration at 0 bar g
PT-113
Calibration at 5 bar g
ÄKTA ready Operating Instructions 28960345 AD
217
14 Reference information
14.6 Checklists