Ordering information for ReadyToProcess columns is found in
. RTP columns packed with Capto and Sepharose 6FF media can
be requested by GE Healthcare even if not listed in ordering list.
ReadyToProcess columns are qualified by efficiency testing and are delivered with indi-
vidual certificates. Testing includes for example analysis of theoretical plates per m
packed bed (N/m) and asymmetry factor (A
s
).
Other columns
Although ReadyToProcess columns are recommended for use with ÄKTA ready, other
columns may be used provided they are capable of withstanding maximum system
pressure. If not, separate pressure safety devices must be used.
WARNING
•
High pressure. The flow rate may under no circumstances
exceed the specified column maximum flow rate. High flows
might affect the packed medium, causing the pressure to ex-
ceed the specified column maximum pressure.
•
Over-pressure. Never block the outlet tubing with, for instance,
stop plugs, since this will create over-pressure and might result
in injury.
Contact GE Healthcare for more information about using products other than ReadyTo-
Process columns with ÄKTA ready.
126
ÄKTA ready Operating Instructions 28960345 AD
9 Column installation
9.1 Columns