14.1
Specifications
General specifications
ÄKTA ready gradient
ÄKTA ready
Property
116 x 80 x 165 cm
100 × 80 ×165 cm
Dimensions (width × depth
× height)
250 kg
230 kg
Weight
•
UNICORN 5.20 or
higher versions of
UNICORN 5
or
•
UNICORN 7.01 or
higher versions of
UNICORN 7
•
UNICORN 5.20 or
higher versions of
UNICORN 5
or
•
UNICORN 7.01 or
higher versions of
UNICORN 7
Control system
1 x 100/120/200-208/220-
230/240 V, ± 10%
1 x 100/120/200-208/220-
230/240 V, ± 10%
AC voltage
50/60Hz
50/60Hz
Frequency
1 kVA
1 kVA
Max. power consumption
IP45, pump IP25
IP45, pump IP25
Ingress protection
5.5 to 10 bar, 50 Nl/min,
oil- and particle-free
5.5 to 10 bar, 50 Nl/min,
oil- and particle-free
Compressed air interface
Recommended operating
conditions
Value
Property
2ºC to 30ºC
Ambient temperature
2ºC to 40ºC
Liquid temperature
950 to 1050 kg/m
3
Fluid density
1.0 to 1.3x water viscosity at actual temperature, max.
2.5 cP
Fluid viscosity
202
ÄKTA ready Operating Instructions 28960345 AD
14 Reference information