UV
Problems with UV can be caused by:
Solution
Problem
Check that the UV cell and sensor windows are clean and
dry. Wipe with soft tissue if necessary.
Insufficient light trans-
mission.
Make sure that the UV cell is pushed all the way in.
Incorrect mounting of
UV cell.
Check that the wash liquid is connected to Inlet 6 and the
test liquid to Inlet 5.
Wrong inlets used.
Make sure that the liquids used are fresh and correctly
prepared (see
).
Wrong composition of
wash liquid and/or test
solution.
Make sure the temperature of the process liquid not is
lower than 10ºC from the ambient temperature.
Condensation of mois-
ture on the flow cell
window.
120
ÄKTA ready Operating Instructions 28960345 AD
8 Component test
8.3 Troubleshooting a failed component test