Sensor specifications
Max. measure-
ment error
1
Acceptance range
Range
Sensor
±0.20 bar g
0 to 5 bar g
-1 to 10 bar g
Pressure
±5% actual value
at flow ≥ 40 l/h
3
7.5 to 510 l/h
0 to 510 l/h
Flow
(High Flow Kit)
2
±10% actual value
at flow < 40 l/h
±5% actual value
at flow ≥ 20 l/h
3 to 175 l/h
0 to 175 l/h
Flow (Low Flow Kit)
±10% actual value
at flow < 20 l/h
±5% full scale
5
0 to 150 mS/cm
0 to 199.99 mS/cm
Conductivity
4
±0.2 pH at calibra-
tion temperature
pH 2 to 12
pH 0 to 14
pH
6
Drift: ±0.025 pH
units/h (20ºC, pH 4)
Flow rate sensitivi-
ty: ±0.1 pH unit
Deviation from lin-
earity ±5%
, 7 , 8
0.01 to 1.0 AU
-7 to 7 AU
UV
2ºC to 40ºC
0 to 47°C
Temperature
1
Max. error, valid within acceptance range under operation conditions.
2
Flow meter cells for High Flow Kit and Low Flow Kit are designed with different flow path ge-
ometries
3
Valid for max. temperature difference of 10 ºC between liquid and ambient temperature.
4
Temperature compensation optional. The standard setting of the cell constant for the sensor
is 12.5 cm-1 and is not to be changed when replacing flow kits.
5
Valid at actual temperature (no temperature compensation).
6
Temperature compensation not available.
7
Cell length 1.8 mm ±0.2 mm.
8
Cell length 1.8 mm ±0.2 mm.
ÄKTA ready Operating Instructions 28960345 AD
205
14 Reference information
14.1 Specifications