4.4
Unpacking
Check for damage
Check the equipment for damage before starting assembly and installation. There are
no loose parts in the crate. All parts are either mounted on the system or located in the
accessory kit box. If any damage is found, document the damage, and contact your local
GE Healthcare representative.
Safety Precautions
WARNING
Personal Protective Equipment (PPE). Whenever packing, unpack-
ing, transporting or moving ÄKTA ready, wear:
•
Protective footwear, preferably with steel lining.
•
Working gloves, protecting against sharp edges.
•
Protective glasses.
WARNING
Because of the significant weight of ÄKTA ready, great care must
be taken not to cause squeezing or crushing injuries during
movement. It is recommended that two persons move the system.
WARNING
At least two persons are required to unpack ÄKTA ready gradient!
Instructions for unpacking
For a full description of how to unpack ÄKTA ready refer to the following unpacking in-
structions:
•
ÄKTA ready Unpacking instructions, article number 28-9944-89, and
•
ÄKTA ready gradient Unpacking instructions, article number 29-0420-82
ÄKTA ready Operating Instructions 28960345 AD
65
4 Installation
4.4 Unpacking