Operating Instructions

iTEMP TMT85

Dual-input temperature transmitter with FOUNDATION
Fieldbus

TM

 protocol

BA00251R/09/EN/16.20

71498847

2020-06-18
Valid as of version

01.02 (device version)

Содержание iTEMP TMT85

Страница 1: ...s Solutions Services Operating Instructions iTEMP TMT85 Dual input temperature transmitter with FOUNDATION FieldbusTM protocol BA00251R 09 EN 16 20 71498847 2020 06 18 Valid as of version 01 02 device version ...

Страница 2: ......

Страница 3: ...ng device and FF functions 30 8 Commissioning 31 8 1 Installation check 31 8 2 Switching on the transmitter 31 8 3 Commissioning the FOUNDATION FieldbusTM interface 31 9 Diagnostics and troubleshooting 37 9 1 Troubleshooting 37 9 2 Status messages 39 9 3 Application errors without messages 44 9 4 Spare parts 45 9 5 Return 46 9 6 Disposal 46 9 7 Software history and overview of compatibility 46 10 ...

Страница 4: ... observed Make sure that you use the right Ex specific documentation for the right device with approval for use in hazardous areas The number of the specific Ex documentation XA is provided on the nameplate If the two numbers on the Ex documentation and the nameplate are identical then you may use this Ex specific documentation 1 3 Symbols 1 3 1 Safety symbols DANGER This symbol alerts you to a da...

Страница 5: ...lant grounding system 1 3 3 Symbols for certain types of information Symbol Meaning Permitted Procedures processes or actions that are permitted Preferred Procedures processes or actions that are preferred Forbidden Procedures processes or actions that are forbidden Tip Indicates additional information Reference to documentation A Reference to page Reference to graphic Notice or individual step to...

Страница 6: ...he technical data on the device and provides an overview of the accessories and other products that can be ordered for the device Brief Operating Instructions KA00252R 09 en Guide that takes you quickly to the 1st measured value The Brief Operating Instructions contain all the essential information from incoming acceptance to initial commissioning The document types listed are available In the Dow...

Страница 7: ... mounting in a terminal head flat face as per DIN EN 50446 It is also possible to mount the device on a DIN rail using the optional DIN rail clip If the device is used in a manner not specified by the manufacturer the protection provided by the device may be impaired The manufacturer is not liable for damage caused by improper or non designated use 2 3 Operational safety Operate the device only if...

Страница 8: ...the EC directives listed in the device specific EC Declaration of Conformity Endress Hauser confirms this by affixing the CE mark to the device 2 5 IT security Our warranty is valid only if the device is installed and used as described in the Operating Instructions The device is equipped with security mechanisms to protect it against any inadvertent changes to the settings IT security measures whi...

Страница 9: ...atures on the delivery note Enter the serial number from the nameplate in the W M Device Viewer www endress com deviceviewer All data relating to the device and an overview of the Technical Documentation supplied with the device are displayed Enter the serial number on the nameplate into the Endress Hauser Operations App or scan the 2 D matrix code QR code on the nameplate with the Endress Hauser ...

Страница 10: ...uidelines The manufacturer confirms that the device is compliant with the relevant guidelines by applying the CE EAC mark 3 4 2 FOUNDATION FieldbusTM certification The temperature transmitter successfully passed all tests and is certified and registered by the Fieldbus Foundation The device meets all the requirements of the following specifications Certified in accordance with FOUNDATION FieldbusT...

Страница 11: ...715 using the DIN rail clip accessory see the Accessories section Information about the conditions that must be present at the installation point such as the ambient temperature degree of protection climate class etc so that the device can be mounted correctly is provided in the Technical data section 47 When using in hazardous areas the limit values of the certificates and approvals must be obser...

Страница 12: ...nting screws 8 Terminal head cover 9 Cable entry Procedure for mounting in a terminal head Item A 1 Open the terminal head cover 8 on the terminal head 2 Guide the connection wires 4 of the insert 3 through the center hole in the head transmitter 5 3 Fit the mounting springs 6 on the mounting screws 7 4 Guide the mounting screws 7 through the side boreholes of the head transmitter and the insert 3...

Страница 13: ...through the lateral bores in the head transmitter 3 3 Screw the head transmitter to the field housing 4 After wiring close the field housing cover 1 again Item C Mounting on DIN rail DIN rail as per IEC 60715 1 Mounting screws with springs 2 Head transmitter 3 Circlips 4 DIN rail clip 5 DIN rail Procedure for mounting on a DIN rail Item C 1 Press the DIN rail clip 4 onto the DIN rail 5 until it en...

Страница 14: ... conditions or special regulations 4 Guide the mounting screws 6 through the lateral bores of the head transmitter 5 5 Position the head transmitter 5 in the terminal head 4 in such a way that the bus cable terminals 1 and 2 point to the cable entry 6 Using a screwdriver screw down the head transmitter 5 in the terminal head 4 7 Guide the connection wires of the insert 3 through the lower cable en...

Страница 15: ... must click securely into place on the head transmitter After mounting securely tighten the terminal head cover The display can be used only with the appropriate terminal heads cover with viewing window e g TA30 from Endress Hauser 4 3 Post installation check After installing the device always run the following final checks Device condition and specifications Notes Is the device undamaged visual i...

Страница 16: ...a mounted head transmitter 1 Open the cable gland and the housing cover on the terminal head or the field housing 2 Feed the cables through the opening in the cable gland 3 Connect the cables as shown in 6 16 If the head transmitter is fitted with spring terminals pay particular attention to the information in the Connecting to spring terminals section 17 4 Tighten the cable gland again and close ...

Страница 17: ... AC between the input and output The following connection combinations are possible when both sensor inputs are assigned Sensor input 1 Sensor input 2 RTD or resistance transmitter two wire RTD or resistance transmitter three wire RTD or resistance transmitter four wire Thermocouple TC voltage transmitter RTD or resistance transmitter two wire RTD or resistance transmitter three wire RTD or resist...

Страница 18: ...able type B several fieldbuses same degree of protection may be operated in one cable No other circuits are permissible in the same cable Practical experience has shown that cable types C and D should not be used due to the lack of shielding since the freedom from interference generally does not meet the requirements described in the standard The electrical data of the fieldbus cable have not been...

Страница 19: ...connected per fieldbus segment However this number is limited under certain conditions explosion protection bus power option field device current consumption A maximum of four field devices can be connected to a spur Shielding and grounding The specifications of the Fieldbus Foundation provided in the Wiring and Installation document must be observed during installation Bus termination The start a...

Страница 20: ...nal cable glands If you replace just one measuring device later on bus communication will have to be interrupted Cable gland or entry Please also observe the general procedure on 16 2 1 3 4 5 6 A0041953 8 Connecting the signal cables and power supply 1 Head transmitter installed in field housing 2 Head transmitter installed in terminal head 3 Terminals for fieldbus communication and power supply 4...

Страница 21: ... 5 mm 1 040 in M20x1 5 NPT 1 2 7 8 16 UNC A B 4 1 2 C 3 5 6 A0043152 9 Connectors for connection to the FOUNDATION FieldbusTM Pin assignment color codes D 7 8 connector A Fieldbus connector 1 1 blue wire FF terminal 2 B Terminal head 2 2 brown wire FF terminal 1 C Connector on housing male 3 Gray wire shield 4 Green yellow wire ground 5 Positioning key Connector technical data Wire cross section 4...

Страница 22: ...sed cable glands with dummy plugs Do not remove the grommet from the cable gland A0024523 10 Connection tips to retain IP67 protection 5 4 Post connection check Device condition and specifications Notes Are the device or cables free from damage visual check Electrical connection Notes Does the supply voltage match the information on the nameplate 9 to 32 VDC Do the cables used meet the required sp...

Страница 23: ...ecifications Notes Has the max length of the fieldbus cable been observed in accordance with the fieldbus specifications 18 Has the max length of the spurs been observed in accordance with the fieldbus specifications Is the fieldbus cable fully shielded and correctly grounded ...

Страница 24: ...re settings for the FOUNDATION Fieldbus TM interface can be made using DIP switches on the rear of the optional display Enabling disabling the simulation mode in the Analog Input function block Switching the hardware write protection on off Switching rotating the display by 180 1 2 A0041955 11 Operation options for the head transmitter 1 Configuration operating programs for operation via FOUNDATIO...

Страница 25: ... Meaning Error message Failure detected An operating error has occurred The measured value is no longer valid The display alternates between the error message and no valid measured value available see the Diagnostics and troubleshooting section 37 Detailed information on the error messages can be found in the Operating Instructions Service mode The device is in service mode e g during a simulation...

Страница 26: ...pts the settings within one second 5 Secure the cover back onto the terminal head or field housing Switching write protection on off Write protection is switched on and off via a DIP switch on the rear of the optional attachable display When write protection is active parameters cannot be modified A lock symbol on the display indicates that write protection is on Write protection prevents any writ...

Страница 27: ... and diagnostic so ware High speed Ethernet HSE 100 Mbit s 32 devices per segment Field controller linking device ENDRESS HAUSER ENDRESS HAUSER linking device linking device H1 31 25 kbit s H1 IEC 61158 2 H1 FISCO Measuring point with installed TMT85 Measuring point with installed TMT85 A0042920 EN 14 System integration via FOUNDATION FieldbusTM The following system connection options are possible...

Страница 28: ...ed by a bus administrator known as the Link Active Scheduler which centrally controls the time sequence of the bus communication process The LAS organizes all the bus activities and sends corresponding commands to the individual field devices Other tasks of the LAS include Recognizing and reporting newly connected devices Logging out devices that are no longer communicating with the fieldbus Maint...

Страница 29: ...he following blocks are available for control tasks PD controller or PID controller For more information please refer to Section 13 ff 7 1 6 Fieldbus based process control With FOUNDATION FieldbusTM field devices can perform simple process control functions themselves and thereby reduce the workload on the superior process control system Here the Link Active Scheduler LAS coordinates data exchange...

Страница 30: ...guration These can be used for configuring both the FF functions and all of the device specific parameters The predefined function blocks allow uniform access to all the network and fieldbus device data The step by step procedure for commissioning the FF functions for the first time is described in detail in the Commissioning section as is the configuration of device specific parameters 33 System ...

Страница 31: ...g this process the following sequence of messages appears on the display Step User interface 1 Display name and the firmware FW and hardware HW version 2 Firm logo 3 Device name and the firmware hardware version and device revision of the head transmitter 4 The sensor configuration is displayed 5 Current measured value or Current status message If the switch on procedure is not successful the rele...

Страница 32: ...ts the individual block to the Out Of Service mode Set to auto mode Resource Sensor Transducer Display Transducer AdvDiagnostic Transducer AI PID and ISEL Sets the individual block to the Auto mode Restart Resource Restarts the device with different options as to which particular parameters should be reset to the factory settings Sensor drift monitoring configuration AdvDiagnostic Transducer Setti...

Страница 33: ...mber Base index Description RS_xxxxxxxxxxx 400 Resource Block TB_S1_xxxxxxxxxxx 500 Transducer Block temperature sensor 1 TB_S2_xxxxxxxxxxx 600 Transducer Block temperature sensor 2 TB_DISP_xxxxxxxxxxx 700 Transducer Block Display local display TB_ADVDIAG_xxxxxxxxxxx 800 Transducer Block Advanced Diagnostic AI_1_xxxxxxxxxxx 900 Analog Input function block 1 AI_2_xxxxxxxxxxx 1000 Analog Input funct...

Страница 34: ...eter to AUTO Configuring the Transducer Blocks The individual Transducer Blocks comprise various parameter groups arranged by device specific functions Temperature sensor 1 Transducer Block TB_S1_xxxxxxxxxxx base index 500 Temperature sensor 2 Transducer Block TB_S2_xxxxxxxxxxx base index 600 Local display functions Transducer Block TB_DISP_xxxxxxxxxxx base index 700 Advanced diagnostics Transduce...

Страница 35: ...es for alarms and warnings using the following parameters HI_HI_LIM limit value for the high alarm HI_LIM limit value for the high warning LO_LIM limit value for the low warning LO_LO_LIM limit value for the low alarm The limit values entered must be within the value range defined in the OUT_SCALE parameter group 18 In addition to the actual limit values the behavior in the event of limit value ov...

Страница 36: ...tor from National Instruments is used to connect the function blocks to the desired control strategy generally graphically and then the time for processing the individual process control functions is specified 20 Once you have specified the active LAS download all the data and parameters to the field device 21 Set the operating mode in the MODE_BLK parameter group TARGET parameter to AUTO This is ...

Страница 37: ...e Replace transmitter Onsite error messages on the display 39 Faulty connection to the fieldbus host system No connection can be made between the fieldbus host system and the device Check the following points Fieldbus connection Check the data cable Fieldbus connector optional Check the pin assignment wiring Fieldbus voltage Check that a min bus voltage of 9 VDC is present at the terminals Permitt...

Страница 38: ...the fieldbus users 33 Analog Input function block Although the operating mode is set to AUTO the status of the AI output value OUT is BAD or UNCERTAIN Check whether an error is pending in the Transducer Block Advanced Diagnostic Transducer Block Adv Diagnostic Actual Status Category and Actual Status Number parameters 39 Parameters cannot be changed or No write access to parameters 1 Parameters th...

Страница 39: ... category The device does not continue measuring with the F status message If a display unit is attached the display alternates between the status message and no valid measured value available Depending on the setting of the Fail Safe Type parameter FSAFE_TYPE the last valid measured value the incorrect measured value or the value configured under Fail Safe Value FSAFE_VALUE is transmitted via the...

Страница 40: ...ted at the sensor terminals Remedy Check the sensor and sensor wiring SV1 SV2 also PV1 PV2 depending on the configuration Transducer_Error Mechanical failure SUBSTATUS Sensor failure F 221 Device status message FF Reference measurement F 221 Local display F221 BLOCK_ERR Other QUALITY BAD Cause of error Internal reference junction defective Remedy Device defective replace SV1 SV2 PV1 PV2 DT Transdu...

Страница 41: ...messages in the relevant Sensor Transducer Block Sensor Transducer Block measured value status default Cause of error remedy Output variables affected M 042 Device status message FF Corrosion M 042 Local display M042 Measured value BLOCK_ERR Other QUALITY UNCERTAIN configurable Cause of error Corrosion detected on the sensor terminals Remedy Check wiring and replace if necessary SV1 SV2 also PV1 P...

Страница 42: ... sensor Remedy Resolve sensor error SV1 SV2 also PV1 PV2 depending on the configuration Transducer_Error No error SUBSTATUS Non specific 9 2 3 Category S diagnostics messages Category No Status messages ACTUAL_STAT US_NUMBER in the Advanced Diagnostics Transducer Block Local display Error messages in the relevant Sensor Transducer Block Sensor Transducer Block measured value status default Cause o...

Страница 43: ...vice initialization C 402 Local display C402 Measured value BLOCK_ERR Power up QUALITY UNCERTAIN Cause of error Device starting initializing Remedy Message is only displayed during power up SV1 SV2 PV1 PV2 DT Transducer_Error Data integrity error SUBSTATUS Non specific C 482 Device status message FF Simulation active C 482 Local display C482 Measured value BLOCK_ERR Other QUALITY UNCERTAIN Cause o...

Страница 44: ...C 10 kΩ 10 kΩ x 15 kΩ 15 kΩ Off ALARM F 041 On F C S M 042 depending on the configuration ALARM F 042 The sensor resistance can affect the resistance data in the table If all the sensor connection cable resistances are increased at the same time the values given in the table are halved The corrosion detection system presumes that this is a slow process with a continuous increase in the resistance ...

Страница 45: ...oint set See Section 13 Offset incorrectly set Check offset Interference via the thermocouple wire welded in the thermowell interference voltage Use a sensor where the thermocouple wire is not welded Sensor connected incorrectly Connect the connecting cables correctly see the Electrical connection section Sensor sensing element defective Check the sensor sensing element Programming Incorrect senso...

Страница 46: ...ware history and overview of compatibility Revision history The firmware version FW on the nameplate and in the Operating Instructions indicates the device release XX YY ZZ example 01 02 01 XX Change to main version No longer compatible The device and Operating Instructions change YY Change to functions and operation Compatible The Operating Instructions change ZZ Fixes and internal changes No cha...

Страница 47: ...ormation on the order code in question is available from your local Endress Hauser sales center or on the product page of the Endress Hauser website www endress com Accessories included in the scope of delivery Printed copy of the Brief Operating Instructions ATEX supplementary documentation ATEX Safety instructions XA Control Drawings CD Mounting material for head transmitter Mounting material fo...

Страница 48: ...Via the Internet https portal endress com webapp applicator Accessories Description Configurator Product Configurator the tool for individual product configuration Up to the minute configuration data Depending on the device Direct input of measuring point specific information such as measuring range or operating language Automatic verification of exclusion criteria Automatic creation of the order ...

Страница 49: ...tion commissioning and operation of the measuring devices All the relevant information is available for every measuring device over the entire life cycle such as the device status device specific documentation spare parts etc The application already contains the data of your Endress Hauser device Endress Hauser also takes care of maintaining and updating the data records W M is available Via the I...

Страница 50: ...cuit compensation of wire resistance possible 0 to 30 Ω With 3 wire and 4 wire connection sensor wire resistance up to max 50 Ω per wire Resistance transmitter Resistance Ω 10 to 400 Ω 10 to 2000 Ω Thermocouples as per standard Designation Measuring range limits IEC 60584 Part 1 Type A W5Re W20Re 30 Type B PtRh30 PtRh6 31 Type E NiCr CuNi 34 Type J Fe CuNi 35 Type K NiCr Ni 36 Type N NiCrSi NiSi 3...

Страница 51: ...ance transmitter four wire Thermocouple TC voltage transmitter 13 2 Output Output signal FOUNDATION FieldbusTM H1 IEC 61158 2 Failure current FDE Fault Disconnection Electronic 0 mA Data transmission rate supported baudrate 31 25 kBit s Signal encoding Manchester II Output data Available values via AI blocks temperature PV temp sensor 1 2 terminal temperature LAS Link Active Scheduler LM Link Mast...

Страница 52: ...Blocks Block description Block Index 1 Execution time macro cycle 500 ms Block category Resource Block 400 Extended Transducer Block Sensor 1 500 Manufacturer specific Transducer Block Sensor 2 600 Manufacturer specific Transducer Block Display 700 Manufacturer specific Transducer Block Adv Diag 800 Manufacturer specific Function Block AI1 900 30 ms Extended Function Block AI2 1000 30 ms Extended ...

Страница 53: ...ped in this Transducer Block Analog Input AI In the AI function block the process variables from the Transducer Blocks are prepared for subsequent automation functions in the control system e g scaling limit value processing PID This function block contains input channel processing proportional integral differential control PID and analog output channel processing The following can be implemented ...

Страница 54: ...erence conditions specified above The measured error data correspond to 2 σ Gaussian distribution The data include non linearities and repeatability Typical Standard Designation Measuring range Typical measured error Resistance thermometer RTD as per standard Digital value 1 IEC 60751 2008 Pt100 1 0 to 200 C 32 to 392 F 0 08 C 0 14 F IEC 60751 2008 Pt1000 4 0 08 K 0 14 F GOST 6651 94 Pt100 9 0 07 ...

Страница 55: ...transmitted via FIELDBUS 2 Maximum measured error for the specified measuring range 3 Deviations from maximum measured error possible due to rounding Measured error for thermocouples TC and voltage transmitters Standard Designation Measuring range Measured error Non repeatabil ity Digital 1 Maximum 2 Based on measured value 3 IEC 60584 1 Type A 30 0 to 2500 C 32 to 4532 F 1 33 C 2 39 F 0 8 C 1 44 ...

Страница 56: ...otal measured error of transmitter at current output Measured error digital Measured error D A Sample calculation with Pt100 measuring range 0 to 200 C 32 to 392 F ambient temperature 25 C 77 F supply voltage 24 V Measured error 0 06 C 0 006 x 200 C 200 C 0 084 C 0 151 F Sample calculation with Pt100 measuring range 0 to 200 C 32 to 392 F ambient temperature 35 C 95 F supply voltage 30 V Measured ...

Страница 57: ...stead of using the standardized sensor curve data Operating influences The measured error data correspond to 2 σ Gaussian distribution Influence of ambient temperature and supply voltage on operation for resistance thermometers RTD and resistance transmitters Designation Standard Ambient temperature Influence per 1 C 1 8 F change Supply voltage Influence per V change Digital 1 Digital 1 Maximum Ba...

Страница 58: ...03 C 0 005 F 0 09 C 0 16 F 0 0045 MV at least 0 03 C 0 005 F Type D 33 ASTM E988 96 0 08 C 0 14 F 0 004 MV at least 0 035 C 0 063 F 0 08 C 0 14 F 0 004 MV at least 0 035 C 0 063 F Type E 34 IEC 60584 1 0 03 C 0 05 F 0 003 MV LRV at least 0 016 C 0 029 F 0 03 C 0 05 F 0 003 MV LRV at least 0 016 C 0 029 F Type J 35 0 02 C 0 04 F 0 0028 MV LRV at least 0 02 C 0 036 F 0 02 C 0 04 F 0 0028 MV LRV at l...

Страница 59: ... F 0 021 measuring span Ni1000 DIN 43760 IPTS 68 0 02 C 0 036 F 0 018 measuring span 0 032 C 0 058 F 0 024 measuring span 0 036 C 0 065 F 0 025 measuring span Cu50 10 OIML R84 2003 GOST 6651 2009 0 053 C 0 095 F 0 013 measuring span 0 084 C 0 151 F 0 016 measuring span 0 094 C 0 169 F 0 016 measuring span Cu100 11 0 027 C 0 049 F 0 019 measuring span 0 042 C 0 076 F 0 026 measuring span 0 047 C 0 ...

Страница 60: ...pan Type L 43 GOST R8 585 2001 0 08 C 0 144 F 0 015 measuring span 0 12 C 0 216 F 0 02 measuring span 0 17 C 0 306 F 0 02 measuring span Voltage transmitter mV 20 to 100 mV 2 µV 0 022 measuring span 3 5 µV 0 03 measuring span 4 7 µV 0 033 measuring span Influence of reference junction Pt100 DIN IEC 60751 Cl B internal reference junction with thermocouples TC 13 5 Environment Ambient temperature ra...

Страница 61: ... Class B equipment Overvoltage category Measurement category II Pollution level Pollution level 2 13 6 Mechanical construction Design dimensions Dimensions in mm in Head transmitter 24 1 0 95 33 1 3 44 1 73 7 0 28 5 0 2 B C A A0007301 17 Version with screw terminals A Spring travel L 5 mm not for US M4 securing screws B Mounting elements for attachable measured value display TID10 C Service interf...

Страница 62: ...M12x1 PA 7 8 FF 40 to 105 C 40 to 221 F TA30A Specification 107 5 4 23 68 5 2 7 28 1 1 78 3 1 15 5 0 6 A0009820 Two cable entries Temperature 50 to 150 C 58 to 302 F without cable gland Material aluminum polyester powder coated Seals silicone Cable entry glands 1 2 NPT and M20x1 5 Color of head blue RAL 5012 Color of cap gray RAL 7035 Weight 330 g 11 64 oz TA30A with display window in cover Specif...

Страница 63: ...lor of aluminum cap gray RAL 7035 Weight Aluminum approx 640 g 22 6 oz Stainless steel approx 2400 g 84 7 oz TA30H with display window in cover Specification 125 4 92 115 4 53 28 1 1 78 3 01 20 5 0 8 A0009831 Flameproof XP version explosion protected captive screw cap with two cable entries Protection class NEMA Type 4x Encl Temperature 50 to 150 C 58 to 302 F for rubber seal without cable gland o...

Страница 64: ...in plated brass contact springs 1 4310 301 AISI Potting PU corresponds to UL94 V0 WEVO PU 403 FP FL fire resistance properties Field housing see specifications 13 7 Certificates and approvals CE mark The product meets the requirements of the harmonized European standards As such it complies with the legal specifications of the EC directives The manufacturer confirms successful testing of the produ...

Страница 65: ...ilable on request The device can also be operated with certified devices of other manufacturers Physical Layer Conformance Test of the Fieldbus FOUNDATIONTM FF 830 FS 2 0 13 8 Supplementary documentation Operating Instructions iTEMP TMT85 BA00251R Brief Operating Instructions iTEMP TMT85 KA00252R Operating Instructions Guideline FOUNDATION Fieldbus Function Blocks BA00062S Supplementary ATEX docum...

Страница 66: ...vice may have other blocks e g several Analog Input function blocks if more than one process variable is available from the field device 1 x Advanced Diagnostic 1 x Display 1 x Sensor 1 FOUNDATION FieldbusTM 1 x Sensor 2 Resource Block 3 x Analog Input 1 x Input Selector Function Blocks Transducer Blocks Function Block 1 x PID Function Block max 6 x Analog Input Function Blocks A0042923 19 TMT85 b...

Страница 67: ...mulation in the Analog Input function block can be disabled or enabled by means of DIP switches on the optional display The WRITE_LOCK parameter shows the status of the hardware write protection The following statuses are possible LOCKED Device data cannot be changed via the FOUNDATION Fieldbus interface NOT LOCKED Device data can be changed via the FOUNDATION Fieldbus interface The BLOCK_ERR para...

Страница 68: ...tem to sort alarms and events User input 1 to 255 Factory setting 0 36 Block Alarm BLOCK_ALM AUTO OOS Displays the current block condition with information on pending configuration hardware or system errors including information on the date and time when the error occurred The block alarm is triggered by the following block errors SIMULATE ACTIVE OUT OF SERVICE If the alarm option is not activated...

Страница 69: ...LT_STATE Read only Current status display of the fault state of the Analog Output and Discrete Output function blocks 54 Check Active FD_CHECK_ACTIVE Read only Indicates whether a diagnostic event of the defined category is currently pending 66 Check Alarm FD_CHECK_ALM AUTO OOS Alarms that are actively transmitted by the device to the fieldbus 58 Check Map FD_CHECK_MAP AUTO OOS Enable or disable d...

Страница 70: ...s for this device 17 Features FEATURES Read only Displays the additional functions supported by the device Display Reports Faultstate Hard W Lock Change Bypass in Auto MVC Report Distribution supported Multi bit Alarm Bit Alarm Support 18 Feature Selection FEATURES_SEL AUTO OOS Use this function to select the additional functions supported by the device 75 FF communication software version FF_COMM...

Страница 71: ...ted in this operating mode OOS Out of Service The block is in the Out of Service mode The execution of the remaining blocks ISEL AI and PID function block is stopped in this operating mode These blocks cannot be set to the AUTO mode The current operating status of the Resource Block is also shown via the RS_STATE parameter 50 Resource Directory RES_DIRECTORY Read only Displays the Resource Directo...

Страница 72: ...ascade SHED_RCAS AUTO OOS Specify the monitoring time for checking the connection between the fieldbus host system and a function block in the RCAS operating mode Once the monitoring time elapses the function block switches from the RCAS operating mode to the operating mode selected in the SHED_OPT parameter Factory setting 640000 1 32 ms 27 Shed Remote Out SHED_ROUT AUTO OOS Specify the monitorin...

Страница 73: ...sducer Blocks The Transducer Blocks of the TMT85 contain all the measurement and device specific parameters All the settings directly associated with the application temperature measurement are made here They form the interface between sensor specific measured value processing and the Analog Input function blocks required for automation A Transducer Block allows the user to influence the input and...

Страница 74: ...ock receives an input value that cannot be evaluated from the Transducer Block a process alarm is generated This process alarm is displayed in the BLOCK_ERR parameter of the Analog Input function block BLOCK_ERR Input Failure The BLOCK_ERR parameter of the Transducer Block 75 displays the device error that produced the input value that could not be evaluated and thus triggered the process alarm in...

Страница 75: ...e blocks thereby enabling faster evaluation Grouping is performed by entering the same numerical value in the STRATEGY parameter of each individual block Factory setting 0 These data are neither checked nor processed by the Transducer Blocks Alert key ALERT_KEY AUTO OOS Use this function to enter the identification number of the plant unit This information can be used by the fieldbus host system t...

Страница 76: ... the active block alarm can be acknowledged in this parameter group The device does not use this parameter to display a process alarm since this is generated in the BLOCK_ALM parameter of the Analog Input function block Transducer Type TRANSDUCER_ TYPE Read only Displays the Transducer Block type Display Sensor Transducer Blocks Custom Sensor Transducer Display Transducer Block Custom Display Tran...

Страница 77: ...e backup function ensures that the process is not interrupted by the failure of an individual sensor and that maximum safety and availability is achieved Sensor drift detection If 2 sensors are connected and the measured values differ by a specified value a diagnostic message is generated in the device The drift detection function can be used to verify the correctness of the measured values and fo...

Страница 78: ...linear curve The complete measuring point device sensor can be adapted to the desired process using linear scaling offset and slope It is necessary to run through the following procedure for this purpose 1 Switch the setting for the SENSOR_CAL_METHOD parameter to user trim standard calibration Then apply the lowest process value to be expected e g 10 C to the sensor of the device This value is the...

Страница 79: ...ing on the sensor element used In addition the upper and lower calculation limits must be entered in the POLY_COEFF_MIN and POLY_COEFF_ MAX parameters The coefficients for the nickel and copper polynomials can be entered with the aid of a wizard in the Sensor 1 and 2 Transducer Blocks Each of the values can be passed onto an AI function block or shown on the display The AI Block and the Display Bl...

Страница 80: ...not compatible Sensor 2 unit and PV unit do not match The sensor 2 unit and the PV unit are not compatible Drift and No Sensor chosen The sensor drift function has been activated but No Sensor has been selected as the sensor type Drift chosen and units do not match The sensor drift function has been activated but the units of the two sensors are not compatible The following table shows all the dev...

Страница 81: ... to SV_1 if SV_2 value T THRESHOLD_VALUE parameter Threshold value THRESHOLD_ VALUE OOS Value for switching in the threshold PV mode Entry in the range from 270 to 2450 C 454 to 4442 F 270 C to 2450 C 454 F to 4442 F Primary value max indicator PV_MAX_INDICATOR AUTO OOS Max indicator for PV is stored in the non volatile memory in intervals of 10 minutes Can be reset Primary value min indicator PV_...

Страница 82: ...erter Calibration highest point CAL_POINT_HI OOS Upper point for linear characteristic calibration this affects offset and slope To be able to write this parameter SENSOR_CAL_METHOD must be set to User trim standard calibration Calibration lowest point CAL_POINT_LO OOS Lower point for linear characteristic calibration this affects offset and slope To be able to write this parameter SENSOR_CAL_METH...

Страница 83: ...omial method Polynom Coeff B POLY_COEFF_B OOS Polynom Coeff C POLY_COEFF_C OOS Polynom Coeff R0 POLY_COEFF_R0 OOS Polynom Nickel Copper Measuring Range Maximum POLY_COEFF_ MAX OOS Upper calculation limit for the RTD polynomial nickel copper linearization Polynom Nickel Copper Measuring Range Minimum POLY_COEFF_ MIN OOS Lower calculation limit for the RTD polynomial nickel copper linearization Devi...

Страница 84: ...rosion monitoring is only possible for RTDs with a 4 wire connection and thermocouples see also 43 Drift detection can be configured with the SENSOR_DRIFT_MONITORING parameter Drift detection can be enabled or disabled If drift detection is enabled and a drift occurs an error or maintenance request is generated A distinction is made between 2 specific modes SENSOR_ DRIFT_MODE In the Overshooting m...

Страница 85: ...her a status is generated if the value set in the SENSOR_DRIFT_LIMIT parameter is undershot Undershooting or overshot Overshooting If Overshooting is selected the corresponding diagnostic event is generated if the limit value is overshot SENSOR_DRIFT_LIMIT In the case of Undershooting the diagnostic event is output if the limit value is undershot Sensor Drift alert value SENSOR_ DRIFT_ALERT_VALUE ...

Страница 86: ...lays the channel that currently has the error with the highest value PREVIOUS_STATUS_CHANNEL indicates the channel where an error last occurred Actual Status Description Previous Status Description PREVIOUS ACTUAL_ STATUS_DESC Read only AUTO OOS Displays the descriptions of the current and previous error status The descriptions can be taken from the description for the Actual Status Number Previou...

Страница 87: ...ured value status BAD UNCERTAIN GOOD can be configured for the individual diagnostic event STATUS_SELECT_ 103 OOS STATUS_SELECT_ 901 OOS STATUS_SELECT_ 902 OOS DIAGNOSIS_ SIMULATION_ ENABLE OOS Enable disable simulation of a diagnostic event DIAGNOSIS_ SIMULATION_ NUMBER AUTO OOS Use this function to select the diagnostic event to be simulated 14 3 9 Transducer Block Display The settings in the Di...

Страница 88: ...ue displayed Maximum 12 letters The value is not shown on the display Decimal places x DISP_VALUE_ X_FORMAT AUTO OOS Use this function to select the number of decimal places displayed Choose between 0 and 4 The option 4 AUTO means that the maximum possible number of decimal places will always appear on the display Possible settings Auto xxxxx xxxx x xxx xx xx xxx Example of parameterization The fo...

Страница 89: ...nput channel processing proportional integral differential control PID and analog output channel processing The configuration of the PID function block depends on the automation task The following can be implemented Basic controls feedforward control cascade control and cascade control with limiting The options available for measured value processing within the PID function block include signal sc...

Страница 90: ...ry Here one bit of the assignment parameters belongs to every event group Default assignments of diagnostic messages to the individual groups are defined in the following table Event weighting Default event category Event source Bit Events in this group Highest weighting Failure F Sensor 31 F041 Sensor break F043 Sensor short circuit Electronics 30 F221 Reference measurement F261 Device electronic...

Страница 91: ...rable area certain diagnostic events can be assigned independently of the event group they belong to They are then removed from the event group and their behavior can be configured individually The following parameters can be assigned to the configurable area of this device 042 Sensor corrosion 103 Drift 901 Ambient temperature too low 902 Ambient temperature too high Bits 16 31 standard range the...

Страница 92: ... be assigned 4 Change the event group bit from 0 to 1 In the configuration systems this is done by ticking the corresponding check box Example The Highest Weighting Configuration Error group contains the events 431 Calibration values and 437 Configuration error These should be categorized as Function check C and no longer as Failure F In the Resource Block search for the Highest Configuration grou...

Страница 93: ...ed for the following events irrespective of the event group they are assigned to in the default setting 042 Sensor corrosion 103 Drift 901 Ambient temperature too low 902 Ambient temperature too high To change the event category the event first has to be assigned to one of the bits 1 to 15 The ConfigArea_1 to ConfigArea_15 parameters in the ADVANCED DIAGNOSTIC ADVDIAG Block are used for this Then ...

Страница 94: ...r a diagnostic event and corrective action In the FD_RECOMMEN_ACT parameter in the Resource Block a description is displayed for the diagnostic event with the highest priority that is currently active This description has the following structure Diagnostic number Diagnostic text with channel ch x troubleshooting recommendations separated by hyphens e g for the diagnostic event sensor break 41 Sens...

Страница 95: ...is 0 The priority can be individually changed for the four assignment parameters The 4 PRI F C S M parameters from the Resource Block are used for this purpose 14 8 2 Suppressing certain events It is possible to suppress certain events during transmission over the bus using a mask While these events are displayed they are not transmitted over the bus This mask can be found in the MASK parameters F...

Страница 96: ...ield devices number 19 M Maximum overall cable length 19 Maximum spur length 19 Mounting location DIN rail DIN rail clip 11 Field housing 11 Terminal head flat face as per DIN 43729 11 N Nameplate 9 Number of field devices 19 O Operation options Local operation 24 Operating tool 24 Overview 24 Overall cable length 19 P Product safety 8 R Return 46 47 S Solid wire 17 Spur length 19 T Terminal assig...

Страница 97: ......

Страница 98: ......

Страница 99: ......

Страница 100: ...www addresses endress com 71498847 71498847 ...