EOX determination
multi X 2500
80
7.1.3
Performing measurements
WARNING
Concentrated sulfuric acid and concentrated acetic acid can cause severe chemical
burns! Wear the appropriate protective clothing when pouring the electrolyte solution!
Observe all instructions and specifications in the safety data sheets!
1.
Fill fresh sulfuric acid into the sulfuric acid container with outlet on the chlorine
module every day (
→
"Changing the sulfuric acid and cleaning the sulfuric acid
container with outlet", p. 101).
2.
Preparing the measuring cell:
Top up the measuring cell with fresh electrolyte solution every day.
Measuring cell "high sensitive": approx. 65 ml
Insert the magnetic stirring rod into the measuring cell. Fit the cover.
Insert the electrodes and connect them electrically.
Connect the measuring gas hose from the sulfuric acid container to the gas
inlet tube using the Teflon screw connection.
Connect the adapter to the extraction hose.
NOTICE! Do not connect the measuring cell "high sensitive" to the intake pump.
Otherwise, electrolyte could get drawn into the pump.
3.
The analyzer and the chlorine module are switched on.
The devices are booted. The status LEDs on the front panel lights up in green
after approximately 30 seconds.
4.
Open the gas supply and set the primary pressure to 4 to 6 bar.
5.
Switch on the PC.
6.
Start the control and analysis software multiWin and log in with user name and
password.
7.
Initialize the analysis system by clicking on [
INITIALIZE ANALYZER
].
The initialization and automatic detection of all connected components will be
executed.
In the window D
EVICE STATUS
any components not yet ready for operation will be
shown in red. The time it takes to preheat the furnace to 950 °C is about 30 to
40 min. During the warm-up phase it is not possible to start a measurement.
If the analyzer is still not ready for measurements after about 40 min., carry out a
fault analysis in accordance with the notes in chapter "Fault removal" on page 106.
8.
Carry out an end point routine to adjust the electrolyte to the optimal operating
range of the measuring cell.
9.
If necessary, test the measuring cell using a direct method.
10.
Providing the samples:
Load the EOX/liquids rack of the autosampler autoX 112 with samples.