multi X 2500
Maintenance and care
99
6.
Remove the pneumatic seal from the housing.
7.
Pull the second PTFE disk from the bottom of the seal.
8.
Carefully separate the black special seal from the ring.
9.
Insert a new special seal into the ring.
10.
Place a PTFE disk (b) into the housing (c).
11.
Place the pneumatic seal (a) into the housing. The hole in
the ring must align with the hole in the housing.
12.
Use the Allen wrench to tighten the connection bushing for
hose no. 11 (see item 4).
13.
Place a second PTFE disk onto the intermediate ring above
the filter.
14.
Place the pneumatic seal onto the auto-protection assembly
and attach it with 4 screws.
The auto-protection assembly can now be reinstalled
into the analyzer.