multi X 2500
Installation and commissioning
43
4
Installation and commissioning
4.1
Installation conditions
See "Technical data" on page 122 for information on the ambient conditions of the
installation location. It may be necessary to control the temperature in the room.
The installation location should have the characteristics of a chemical laboratory. It
must meet the following conditions:
Atmosphere free from hydrocarbons, halogens, sulfur compounds and nitrogen
oxides
Atmosphere with low dust levels
No vibrations
Smoking must be prohibited in the room where the multi X 2500 is operated
The site for installing the multi X 2500 must meet the following requirements:
Free from caustic vapors in close proximity of the multi X 2500
Connections and assemblies of the device may corrode.
Free from draft air; do not install the device close to windows or doors
Away from electromagnetic sources of interference
No direct sunlight and away from radiant heaters
The front door and air vents must not be obstructed by other equipment or
furnishings
Keep a safety distance of at least 20 cm from the rear of the device to other
equipment or walls
4.2
Energy supply
Caution
The multi X 2500 must only be connected to a power outlet which is properly
grounded in accordance with the voltage specifications on the type plate!
The installation of the electrical equipment of the laboratory must comply with the
standard DIN VDE 0100. After the connection point an electrical current in accordance
with the standard IEC 38 must be available.
Electrical connection data, see "Technical data", p. 122 ff.
Ambient conditions
Laboratory conditions
Requirements with
regard to the installation
location