AOX determination
multi X 2500
62
Create an analysis sequence.
Enter the sample ID for all samples in the field N
AME
.
Confirm the entry with [OK].
Click the [S
TART MEASUREMENT
] button.
The prepared analysis sequence is processed.
14.
Follow the prompts in the control and analysis software multiWin and insert the
one sample after the other through the oxygen sluice into the combustion tube.
Confirm each inserted sample with [OK].
The loaded activated carbon is combusted in the oxygen stream and turned
into hydrogen halide, carbon dioxide and water. After drying the pyrolysis gas
the halide is determined in micro-coulometrically in the chlorine module.
6.2.4
Carrying out measurements with the autosampler autoX 36 / autoX 36d
WARNING
Concentrated sulfuric acid can cause severe chemical burns! Wear appropriate
protective clothing when working at the sulfuric acid container. Observe all
specifications in the safety data sheet!
1.
Connect the autosampler to the "sampler" connection on the rear panel of the
multi X 2500 using the interface cable.
2.
Fill the sulfuric acid container (vertical) with fresh sulfuric acid every day.
If necessary, remove any used quartz containers from the open combustion tube.
3.
When using the autosampler autoX 36d (with ejection function):
Use tweezers to place the quartz crucible into the combustion
tube.
The quartz crucible prevents the quartz wool to form lumps
when it directly touches the hot activated carbon. This would
prevent it from filtering ashes and soot from the measuring
gas stream.
4.
Remove the top cover of the analyzer and insert the autosampler without rack into
the opening.
5.
Preparing the measuring cell:
Fill the measuring cell with fresh electrolyte solution every day.
Measuring cell "sensitive": 15 to 20 ml
Insert the magnetic stirring rod into the measuring cell. Attach the cover and
tighten it using the three knurled head screw.
Insert the combined electrode and connect it electrically. Electrically connect
the measuring cell.