multi X 2500
Safety instructions
11
2.4
Safety instructions - operation
2.4.1
General
The operator of the analyzer must make sure before each commissioning that the
condition of the analyzers including the safety equipment is sound. This applies in
particular after each modification or extension of the analyzer or its repair.
Observe the following notes:
The analyzer must only be operated if all protective equipment (e.g. covers, drip
pans for chemicals, side walls and doors) are in place, properly installed and fully
operational.
The sound condition of the protection and safety equipment must be checked
regularly. Any defects must be corrected as soon as they occur. Protective and
safety equipment must never be removed, modified or switched off during
operation.
Free access to the device switch on the rear panel of device must always be ensured
during operation.
The ventilation equipment on the multi X 2500 and the chlorine module must be in
good working condition. Covered ventilation grilles or slits etc. may cause the
device to break down or may cause damage to it.
The furnace operates with temperatures of up to 1100 °C. The hot components
must not be touched during or directly after the operation of the analyzer.
Caution when handing quartz glass and glass parts. Risk of broken glass and
therefore risk of injury!
Prevent any ingress of liquids into the interior of the device. Liquids may cause a
short circuit inside the housing.
Keep all combustible materials away from the analyzer.
The multi X 2500 may not be operated in an explosive environment.
Smoking or naked flames are prohibited in the room where the multi X 2500 is
operated!
The operating personnel has to be familiar with the location of the fire-fighting
equipment in the operating room of the analyzer.
2.4.2
Safety instructions - electrical equipment
Contact with live components may cause death, serious injury or painful electrical
shock. Life-threatening electrical voltages may occur in the right part of the
multi X 2500 behind the control electronics cover.
Protection against
explosion and fire