Function and layout of the multi X 2500
multi X 2500
42
Fig. 39 Pre-combustion adapter
1 Pneumatic box
2 Interface cable
3 Connection to oxygen supply
4 Pre-combustion adapter
5 Clamping device for the sample
6 Pneumatic unit with gas hose
7 Gas connection on the multi X 2500
The pre-combustion adapter is attached to the open combustion tube instead of the
oxygen sluice. In addition to the oxygen connection it has a pneumatic unit which is
used to move the clamping device for the sample downward. The pre-combustion
adapter is connected to the multi X 2500 via the pneumatic box. It is then possible to
use the multiWin control software to determine for how long the sample is supposed
to be pre-dried in the upper part of the combustion tube clamped in between the glass
tube and the clamping device. After the preset time has elapsed, a gas blast pushes the
clamping device downward. The sample drops into the hot combustion tube where it is
combusted in the oxygen stream.