Operation
multi X 2500
50
5.2
Switching off the analyzer
WARNING
Concentrated sulfuric acid can cause severe chemical burns! Wear protective clothing
when working at the sulfuric acid container. Observe all specifications in the safety
data sheet!
1.
Exit the multiWin control and analysis software. Allow the device to cool down.
NOTICE! Wait for a moment before switching off the basic device and the chlorine
module. Otherwise, the device may overheat when the ventilation is switched off
with the device.
2.
Use the device switch to switch off the basic device and the chlorine module.
3.
Remove the electrode from the measuring cell and store it in accordance with the
applicable regulations (
→
"combined electrode", p. 93).
4.
Empty the measuring cell and clean it (
→
"Servicing the measuring cell", p. 92).
5.
After the device has cooled down, remove the sulfuric acid container from the basic
device and clean it.
NOTICE
Do not cut the gas supply before the sulfuric acid has been drained or the device has
cooled down. A vacuum may build up when the device cools down which may draw
sulfuric acid into the gas transfer line. As long as the gas supply is open, the argon
safety bypass on the auto-protection assembly prevents this from happening.
1.
Exit the multiWin control and analysis software.
2.
Remove the gas transfer line from the sulfuric acid container in the chlorine
module and drain the sulfuric acid from the sulfuric acid container (
→
the sulfuric acid and cleaning the sulfuric acid container with outlet", p. 101).
WARNING! Sulfuric acid can cause severe chemical burns. The grinding of the
sulfuric acid container and the hose which is immersed in the sulfuric acid container
may contain residual acid.
3.
Allow the analyzer to cool down. After that, switch off the multi X 2500 and the
chlorine module using the mains switch. Cut the gas supply.
4.
Remove the electrode from the measuring cell and store it in accordance with the
applicable regulations (
→
"Clean and store the electrode", p. 105).
5.
Empty the measuring cell and clean it (
→
"Maintaining the measuring cell",
p. 104).
Standard version
Optional versions