multi X 2500
Fault removal
107
2
Communication error: Serial interface not available!
Cause
Remedy
Communication problems
Disconnect USB connection between basic device
and PC and re-connect after approx. 10 s
Initialize
3
Communication error: Serial interface not available!
Cause
Remedy
Communication problems
Disconnect USB connection between basic device
and PC and re-connect after approx. 10 s
Initialize
7
Operating system error: Not authorized accesses!
Cause
Remedy
Undefined termination of multiWin
Exit multiWin
Detach USB cable and re-connect after approx.
10 s
Restart operating system (PC)
Restart firmware (basic device)
Restart multiWin
12
Signal echo received, check interface selection
Cause
Remedy
Incorrect port selected
Check port selection
14
Data transfer interrupted
Cause
Remedy
No data transfer for 10 s
Initialize
17
Wrong interface protocol ID
Cause
Remedy
Fault after update
(program versions of firmware multiWin don't match)
Update required
20
Timeout InitEnd
Cause
Remedy
Timeout during initialization
Initialize
21
Timeout StatusBusy
Cause
Remedy
Timeout during operation
(device not ready to measure)
Acknowledge message
Initialize
22
Timeout End
Cause
Remedy
Timeout when exiting multiWin
Acknowledge message
Initialize
23
Timeout StopEnd
Cause
Remedy
Timeout during measurement cancellation
Acknowledge message
Initialize
Содержание Analytik Jena Multi X 2500