Fault removal
multi X 2500
112
Furnace temperature is outside tolerance limits or target temperature is not reached
Cause
Solution
Temperature controller faulty
Electronics error
Report to customer service
Process gases (inlet flow) are not present
Cause
Solution
Gas supply not connected
Primary gas pressure too low
Gas supply leaking
No method loaded
Gas box faulty
Connecting the gas supply
Set primary gas pressure at the delivery point to
between 4 and 6 bar
Check gas supply
Load method
Report to customer service
Target flow at the outlet to the detector module too low
Cause
Solution
Connection between hose, angled adapter and
combustion tube not correct
Pneumatic seal in the auto-protection assembly does
not seal combustion tube
Septum sits incorrectly at the inlet of the combustion
tube or is leaking
Connection transfer line to the auto-protection
assembly leaking
In the horizontal operating mode transition coupling
tube to ABD leaking
Check connection and ensure correct fit at the
connection points
Check argon supply, open the pneumatic seal with the
toggle switch
Check position of the septum, insert new septum if
necessary
Check connection
Check sealing of the coupling element
check alignment between combustion tube, coupling,
ABD, hand-tighten the connection
Gas escapes from pneumatic seal (audible hiss)
Cause
Solution
Screw connection of the gas connection for hose no. 8 is
loose
Pneumatic seal faulty
Tighten screw connection and lock nut
Replace pneumatic seal (
→
chapter 10.4.3, p. 96)
Auto-protection assembly does not heat up
Cause
Solution
Valve assembly not connected
Heating or temperature controller defective
Plug in the connector (3 in Fig. 6, p. 21)
Inform customer service
Auto-injector is not recognized
Cause
Solution
Auto-injector and autoX 112 are switched on at the
same time
Switch off the autosampler autoX 112