POX determination
multi X 2500
84
WARNING
Organically bound halogens must only be purged with argon! Otherwise, there is a risk
of explosion! The purging equipment must only be connected to the "ABD OUT"
terminal on the rear panel of the basic device! Before starting the measurement, verify
that the purging equipment is connected to the right terminal!
WARNING
Concentrated sulfuric acid can cause severe chemical burns! Wear appropriate
protective clothing when working at the sulfuric acid container. Observe all
specifications in the safety data sheet!
CAUTION
Risk of burns when removing the oxygen sluice! The oxygen sluice will get very hot
during the operation of the combustion furnace! Wear protective gloves when
removing the oxygen sluice!
1.
Fill the sulfuric acid container (vertical) with fresh sulfuric acid every day.
If necessary, remove any used quartz containers from the open combustion tube
and clean it.
2.
Place the top cover on top of the analyzer.
3.
Preparing the measuring cell:
Fill the measuring cell with fresh electrolyte solution every day.
Measuring cell "sensitive":
15 to 20 ml
Insert the magnetic stirring rod into the measuring cell. Attach the cover and
tighten it using the three knurled head screw.
Insert the combined electrode and connect it electrically. Electrically connect
the measuring cell.
Connect the gas transfer hose from the sulfuric acid container in the basic
device to the combined electrode. Connect the hose for the adsorption tube
(hose no. 5) via the olive-tip suction adapter to the measuring cell "sensitive".
4.
The analyzer and the chlorine module are switched on.
The devices are booted. The status LEDs on the front panel lights up in green
after approximately 30 seconds.
5.
Open the gas supply and set the primary pressure to 4 to 6 bar.
6.
Switch on the PC.
7.
Start the control and analysis software multiWin and log in with user name and
password.
8.
Initialize the analysis system by clicking on [
INITIALIZE ANALYZER
].
The initialization and automatic detection of all connected components will be
executed.