multi X 2500
Function and layout of the multi X 2500
39
The autosamplers autoX 36 and autoX 36d were specifically conceived for the
multi X 2500. The autosamplers are described in this operating manual. The manual
boat feeder (MBD) is also described in this manual. When using any other sample
introduction module, refer to the respective operating instructions.
3.6.1
autoX 36 / autoX 36d
The autosamplers autoX 36 and autoX 36d are used for small AOX sample series in
vertical operating mode. They can transfer up to 36 samples into the open combustion
tube.
autoX 36
: Autosampler used to automatically introduce a max. of 36 quartz
containers (18 x 6 mm, 18 x 8 mm) with activated carbon which has been enriched
using the column or the batch method.
autoX 36d: Autosampler with ejection function for direct dosing of activated carbon
from a maximum of 36 columns (18 x 6 mm). It can be employed for samples
which were enriched using the column method.
Both autosamplers are connected to the multi X 2500 by means of the interface cable.
The multiWin control software automatically detects autosamplers and sample racks
and controls the introduction of samples. A cover ensures that cross-contamination
and environmental influences can be excluded. On the autoX 36d autosampler the
cover is also used to push down the quartz containers during the ejection process. The
floating arrangement allows the autosampler to be inserted in the right position of the
oxygen sluice in the cutout of the basic device.
Fig. 34 Autosampler autoX 36
1 Interface cable
2 Drop position
3 Cutout for the oxygen sluice
4 Floating bearing arrangement
5 Lug for correct positioning of the drop position
6 Rack