Installation and commissioning
multi X 2500
46
4.5
Setting up and connecting the analyzer
CAUTION
The multi X 2500 must only be set up, assembled and installed by the customer service
department of Analytik Jena or trained specialist personnel authorized by Analytik
Jena!
Any unauthorized intervention on the multi X 2500 can endanger the user and the
operational safety of the equipment and limits or completely invalidates any warranty
claims.
NOTICE
Retain the transport packaging! Return transport for service must be in the original
packaging. This alone prevents transport damage.
The basic module is unpacked and assembled by the customer service department of
Analytik Jena AG or authorized and trained specialist personnel.
Check when unpacking the device for completeness and soundness of the delivery in
accordance with the packing list included. After assembly the customer service tests
the functionality of the multi X 2500 and other system components and documents
the test.
NOTICE
Always switch off all components of the multi X 2500 before electrically connecting
them! Connecting or disconnecting electrical contacts may damage the sensitive
electronic components of the multi X 2500 and detector modules.
NOTICE
Condensed water and temperature differences may damage the electronic components
of the multi X 2500.
Allow the multi X 2500 to acclimatize in the operating room for one hour before
putting the device into service after transport and storage.
Connecting the
multi X 2500