multi X 2500
EOX determination
81
Alternatively, keep the samples at hand for introducing them manually with a
syringe or an auto-injector.
11.
Create a new method under M
ETHOD
M
ETHOD NEW
and release it.
12.
Activate the new method or an already existing method using the menu command
M
ETHOD
M
ETHOD
–
ACTIVATE
.
13.
Starting the measurement:
Select S
TART
S
TART
-A
NALYSIS
.
Select an analysis group or create a new one.
Create an analysis sequence.
Enter the sample ID for all samples in the field N
AME
.
Confirm the entry with [OK].
Click the [S
TART MEASUREMENT
] button.
The prepared analysis sequence is processed.
For manual sample introduction, follow the prompts in the control and analysis
software multiWin and introduce the samples one after the other.
Содержание Analytik Jena Multi X 2500
Страница 1: ...Operating Manual multi X 2500 ...