AOX determination
multi X 2500
70
7.
Attach a funnel to the sulfuric acid container. Fill 15 ml of
concentrated sulfuric acid into the sulfuric acid container.
8.
Remove the funnel.
9.
Fit the safety attachment (7) to the sulfuric acid container
and attach the safety clip.
10.
Connect the measuring gas hose (8) to the electrode (see
below).
Inserting the measuring cell:
NOTICE
Risk of damaging the intake pump of the chlorine module!
Make sure to properly connect the gas supply in the measuring cell when using the
multi-purpose combustion tube. Never connect the measuring cell to the intake pump
via the olive-tip suction adapter and the hose. Otherwise, electrolyte will be drawn into
the pump unit.
The measuring cells "sensitive" and "high concentration" are inserted to the chlorine
module as follows:
1.
Insert the measuring cell "sensitive" with lid into the chlorine module. If used, insert
the measuring cell "high concentration" into the chlorine module
(TX determination).