Fault removal
multi X 2500
108
50
Firmware reset
Cause
Remedy
Internal computer (firmware) restarted
Acknowledge message
Initialize
61
62
64
65
66
67
Incomplete command from PC
Command from PC without STX
Command from PC: CRC error
Command from PC is invalid command
Command from PC: Invalid MEASURING command
Command MTXT from PC is missing
Cause
Remedy
Communication error
Acknowledge message
Initialize
130
Cl sensor: no connection
Cause
Remedy
Communication faulty after detector was detected
during initialization
Acknowledge message
Initialize
131
Cl sensor: wrong command structure
Cause
Remedy
Error in communication with chlorine module
Acknowledge message
Switch chlorine module on/off
Initialize
132
Cl sensor: Indication error
Cause
Remedy
Indicator value after titration start out of range
(measurement cannot start)
Acknowledge message
Initialize
Execution of the end point routine
Check status of measuring cell via S
YSTEM
COMPONENT TEST
C
HLORINE
133
Cl sensor: wrong cell type
Cause
Remedy
No initialization after cell change
Initialize
134
Cl sensor: wrong status
Cause
Remedy
Communication error
Acknowledge message
Initialize
Check titration cell status via
menu item S
YSTEM
COMPONENT TEST
tab
C
HLORINE
135
Cl sensor: wrong version
Cause
Remedy
Transmission error
Acknowledge message
Initialize
Check status of titration cell via
menu item