multi X 2500
Maintenance and care
101
8.
Pull hose no. 3 (1) and hose no. 4 (2) from the angled
adapters at the combustion tube. If used, remove the flame
sensor from the connection (3).
9.
Carefully remove the combustion tube from the furnace.
10.
Check the combustion tube for excessive crystallization,
cracks and blow-outs. Only reuse intact combustion tubes.
11.
Clean the combustion tube with detergent and a brush or cellulose, if necessary.
The combustion tube can also be cleaned by burning off deposits in a muffle
furnace or with a suitable burner flame.
Refer to "Inserting the multi-purpose combustion tube" S. 67 for information on the
reinstallation of the multi-purpose combustion tube.
10.4.5
Changing the sulfuric acid and cleaning the sulfuric acid container with outlet
WARNING
Concentrated sulfuric acid can cause severe chemical burns! Wear protective clothing
when working at the sulfuric acid container. Observe all specifications in the safety
data sheet!
The sulfuric acid absorbs the water that is produced during the combustion process. If
the acid content falls below 85 % the sulfuric acid can no longer sufficiently dry the
reaction gas. As a result, the measured AOX contents will be too low and widely
scattered. For this reason, the sulfuric acid must be replaced once per day. High sample
throughput may require more frequent replacement.