68
MD 610_2d 11/2019
Chlorine HR, free
with Powder Pack
plastic vial (type
3)
10 mm
0.1 – 8 mg/l Cl
2
1.
Fill a clean vial (10 mm Ø) with
5 ml of the water
sample
, closetightly with the cap.
2.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
3.
Press
ZERO
key.
4.
Removethe vial from the sample chamber.
5.
Add the contents of
two Chlorine Free-DPD / F10
Powder Pack
straight from the foil into the water
sample.
6.
Close the vial tightly with the cap and invert several
times to mix the contents (20 sec.).
7.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
8.
Pressthe
TEST
key.
The result is shown in the display in mg/l free Chlo-
rine.
Notes:
See page 47
1
2
3
4
1
1
1
10 ml
5 ml
3
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Содержание Lovibond MD 610
Страница 14: ...8 MD 610_2d 11 2019...
Страница 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Страница 87: ...81 MD 610_2d 11 2019...
Страница 95: ...89 MD 610_2d 11 2019...
Страница 279: ...273 MD 610_2d 11 2019 Notes 1 SO3 Na2 SO3 Reagent Formof reagent Quantity Order No SULFITE LR Tablet 100 518020BT...
Страница 306: ...300 MD 610_2d 11 2019...
Страница 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Страница 310: ...304 MD 610_2d 11 2019...
Страница 362: ...356 MD 610_2d 11 2019...
Страница 363: ...357 MD 610_2d 11 2019 Part 3 Enclosure...