206
MD 610_2d 11/2019
Ozone, in the presence of Chlorine
with Tablet
0.02 – 2 mg/l O
3
1.
Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
, closetightly with the cap.
2.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
3.
Press
ZERO
key.
4.
Removethe vial from the sample chamber and
empty
it, leaving a few drops remaining in the vial
.
5. Add
one DPD No. 1 tablet
and
one DPDNo. 3 tablet
straight from the foil and crushthe tabletsusing a clean
stirring rod.
6.
Add water sample to the 10 ml mark.
7.
Close the vial tightly with the cap and swirl several
times until the tablets are dissolved.
8.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
9.
Press
TEST
key.
Wait for a
reaction period of 2 minutes
.
After the reaction period is finished the measurement
starts automatically.
10. Removethe vial from the sample chamber, empty the
vial, rinse vial and cap several times.
11.
Fill a secondcleanvial with 10 ml of water sample.
12. Add
one GLYCINEtablet
straight from the foil and
crush the tablet using a clean stirring rod.
prepare Zero
press ZERO
Zero accepted
prepare T1
press TEST
Countdown
2:00
3
0
0
Содержание Lovibond MD 610
Страница 14: ...8 MD 610_2d 11 2019...
Страница 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Страница 87: ...81 MD 610_2d 11 2019...
Страница 95: ...89 MD 610_2d 11 2019...
Страница 279: ...273 MD 610_2d 11 2019 Notes 1 SO3 Na2 SO3 Reagent Formof reagent Quantity Order No SULFITE LR Tablet 100 518020BT...
Страница 306: ...300 MD 610_2d 11 2019...
Страница 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Страница 310: ...304 MD 610_2d 11 2019...
Страница 362: ...356 MD 610_2d 11 2019...
Страница 363: ...357 MD 610_2d 11 2019 Part 3 Enclosure...