31
MD 610_2d 11/2019
Notes:
1. Extremelybasic or acidic water samplesshould be adjusted with 0.5 mol/l (1 N) Sulfuric
acid solution or 1 mol/l (1 N) Sodium hydroxide solution to pH 7.
2. Interferences:
Interfering substance
Interference levels and treatments
Calcium
greater than 1000 mg/l CaCO
3
Iron
Interferes at all levels. Correct as follows:
a) determine the concentration of iron present in the
sample by performing a total Iron test
b) add the same iron concentration as determined
to the deionised water (step 1).
The interference will be blanked out successfully.
Magnesium
greater than 6000 mg/l CaCO
3
Nitrate
greater than 100 mg/l NO
3
-N
Nitrite
greater than 12 mg/l NO
2
-N
Phosphate
greater than 100 mg/l PO
4
-P
Sulfate
greater than 300 mg/l SO
4
Sulfide
intensifies the colour
Glycine, Hydrazine,
Colour, Turbidity
Lesscommon interferences such as Hydrazine and
Glycine will causeintensified colours in the prepared
sample. Turbidity and colour will give erroneous high
values. Sampleswith severeinterferences require
distillation.
3.
N
NH
4
NH
3
Reagent
Form of reagent/Quantity
Order-No.
Set
VARIO Ammonia Salicylate F10
VARIO Ammonia Cyanurate F10
Powder Pack / per 100 PP
535500
Содержание Lovibond MD 610
Страница 14: ...8 MD 610_2d 11 2019...
Страница 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Страница 87: ...81 MD 610_2d 11 2019...
Страница 95: ...89 MD 610_2d 11 2019...
Страница 279: ...273 MD 610_2d 11 2019 Notes 1 SO3 Na2 SO3 Reagent Formof reagent Quantity Order No SULFITE LR Tablet 100 518020BT...
Страница 306: ...300 MD 610_2d 11 2019...
Страница 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Страница 310: ...304 MD 610_2d 11 2019...
Страница 362: ...356 MD 610_2d 11 2019...
Страница 363: ...357 MD 610_2d 11 2019 Part 3 Enclosure...