44
MD 610_2d 11/2019
Chloride
with Liquid Reagent
0.5 – 20 mg/l Cl
–
1.
Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
, closetightly with the cap.
2.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
3.
Press
ZERO
key.
4.
Removethe vial from the sample chamber.
5.
Fill the vial with drops of the same size by holding the
bottle vertically and squeezeslowly:
20 drops KS251 (Chloride Reagent A)
6.
Close the vial tightly with the cap and invert several
times to mix the contents.
7.
Fill the vial with drops of the same size by holding the
bottle vertically and squeezeslowly:
20 drops KS253 (Chloride Reagent B)
8.
Close the vial tightly with the cap and invert several
times to mix the contents.
8.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
9.
Press
TEST
key.
Wait for a
reaction period of 5 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Chloride.
9
2
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
5:00
Содержание Lovibond MD 610
Страница 14: ...8 MD 610_2d 11 2019...
Страница 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Страница 87: ...81 MD 610_2d 11 2019...
Страница 95: ...89 MD 610_2d 11 2019...
Страница 279: ...273 MD 610_2d 11 2019 Notes 1 SO3 Na2 SO3 Reagent Formof reagent Quantity Order No SULFITE LR Tablet 100 518020BT...
Страница 306: ...300 MD 610_2d 11 2019...
Страница 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Страница 310: ...304 MD 610_2d 11 2019...
Страница 362: ...356 MD 610_2d 11 2019...
Страница 363: ...357 MD 610_2d 11 2019 Part 3 Enclosure...