33
MD 610_2d 11/2019
Notes:
1. Strong alkaline or acidic water samplesmust be adjusted to approx. pH 7 before analysis
(use 1 mol/l Hydrochloric acid resp. 1 mol/l Sodium hydroxide).
2. Iron interferes with the test. The interferences will be eliminated as follows:
Determine the amount of total iron present in the water sample. To produce the blank
add an iron standard solution with the same iron concentration to the vial (point 1)
instead of deionised water
3. Conversion:
mg/l NH
4
= mg/l N x 1.29
mg/l NH
3
= mg/l N x 1.22
4.
N
NH
4
NH
3
Reagent
Form of reagent/Quantity
Order-No.
Set
VARIO Ammonia Salicylate F5
VARIO Ammonia Cyanurate F5
VARIO Am Diluent Reagent LR
VARIO deionised water
Set
Powder Pack / 50
Powder Pack / 50
Reaction tube / 50
100 ml
535600
Содержание Lovibond MD 610
Страница 14: ...8 MD 610_2d 11 2019...
Страница 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Страница 87: ...81 MD 610_2d 11 2019...
Страница 95: ...89 MD 610_2d 11 2019...
Страница 279: ...273 MD 610_2d 11 2019 Notes 1 SO3 Na2 SO3 Reagent Formof reagent Quantity Order No SULFITE LR Tablet 100 518020BT...
Страница 306: ...300 MD 610_2d 11 2019...
Страница 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Страница 310: ...304 MD 610_2d 11 2019...
Страница 362: ...356 MD 610_2d 11 2019...
Страница 363: ...357 MD 610_2d 11 2019 Part 3 Enclosure...