130
MD 610_2d 11/2019
Hardness, Calcium
with Tablet
50 – 900 mg/l CaCO
3
1.
Filla cleanvial (24 mm Ø)with
10 ml deionised water
.
2.
Add
one CALCHECKtablet
straight from the foil to
the deionised water and crush the tablet using a clean
stirring rod.
3.
Closethe vial tightly with the cap and swirl severaltimes
until the tablet is dissolved.
4.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
5.
Press
ZERO
key.
Wait for a
reaction period of 2 minutes
.
After the reaction period is finished the measurement
starts automatically.
6.
Removethe vial from the sample chamber.
7. Add
2 ml of the water sample
to the prepared vial.
Caution: Vial is filled up to the top! (Note 4)
8.
Closethe vial tightly with the cap and swirl severaltimes
(5x) to mix the contents.
9.
Placethe vial in the sample chamber making sure that
the
marks are aligned.
10. Press
TEST
key.
The result is shown in the display as Calcium Hardness.
Zero accepted
prepare Test
press TEST
prepare Zero
press ZERO
Countdown
2:00
9
0
1
Содержание Lovibond MD 610
Страница 14: ...8 MD 610_2d 11 2019...
Страница 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Страница 87: ...81 MD 610_2d 11 2019...
Страница 95: ...89 MD 610_2d 11 2019...
Страница 279: ...273 MD 610_2d 11 2019 Notes 1 SO3 Na2 SO3 Reagent Formof reagent Quantity Order No SULFITE LR Tablet 100 518020BT...
Страница 306: ...300 MD 610_2d 11 2019...
Страница 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Страница 310: ...304 MD 610_2d 11 2019...
Страница 362: ...356 MD 610_2d 11 2019...
Страница 363: ...357 MD 610_2d 11 2019 Part 3 Enclosure...