99
MD 610_2d 11/2019
Notes:
1. Run samplesand blanks with the same batch of vials. The blank is stable when stored in
the dark and can be used for further measurementswith vials of the same batch.
2. Do not place the hot vials in the sample chamber. Cool the vials to room temperature for
final measurements.
3. Suspendedsolids in the vial lead to incorrect measurements. For this reason it is
important to place the vials carefully in the sample chamber. Theprecipitate at the
bottom of the sample should be not suspended.
4. Clean the outside of the vials with a towel. Finger prints or other marks will be removed.
5. Samplescan be measured when the Chloride content does not exceed10000 mg/l.
6. In exceptional cases,compounds contained in the water cannot be oxidized adequately,
so results may be lower than reference methods.
7. For samplesunder 1 g/l COD it is recommended to repeat the test with the test kit for
COD MR or for samplesunder 0,1 g/l COD with the tube test COD LR.
Reagent
Form of reagent/Quantity
Order-No.
CSBVARIO HR 200 - 15000 mg/l
1 Set (25 tests)
2420722
Содержание Lovibond MD 610
Страница 14: ...8 MD 610_2d 11 2019...
Страница 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Страница 87: ...81 MD 610_2d 11 2019...
Страница 95: ...89 MD 610_2d 11 2019...
Страница 279: ...273 MD 610_2d 11 2019 Notes 1 SO3 Na2 SO3 Reagent Formof reagent Quantity Order No SULFITE LR Tablet 100 518020BT...
Страница 306: ...300 MD 610_2d 11 2019...
Страница 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Страница 310: ...304 MD 610_2d 11 2019...
Страница 362: ...356 MD 610_2d 11 2019...
Страница 363: ...357 MD 610_2d 11 2019 Part 3 Enclosure...