30
MD 610_2d 11/2019
Ammonia
with Vario Powder Pack
0.01 – 0.8 mg/l N
Use two clean vials (24 mm Ø) and mark one as blank for
zeroing.
1.
Fill a clean vial (24 mm Ø) with
10 ml of deionised
water
(this is the blank).
2.
Fill the other clean vial (24 mm Ø) with
10 ml of the
water sample
(this is the sample).
3.
Add the contents of
one Vario Ammonia Salicylate
F10 Powder Pack
straight from the foil to each vial.
4. Close the vials with
the caps and shake to mix the
contents.
5. Press
[
]
key.
Wait for a
reaction period of 3 minutes
.
After the reaction period is finished proceed as follows:
6. Add the contents of
one Vario Ammonia Cyanurate
F10 Powder Pack
straight from the foil to each sam-
ple.
7. Close the vials tightly with
the caps and shake thor-
oughly until the reagent is dissolvedcompletely.
8. Press
[
]
key.
Wait for a
reaction period of 15 minutes
.
After the reaction period is finished proceed as follows:
9.
Placethe vial (the blank) in the samplechamber making
sure that the
marks are aligned.
10. Press
ZERO
key.
11. Removethe vial from the sample chamber.
12. Place the vial (the sample) in the sample chamber
making sure that the
marks are aligned.
13. Press
TEST
key.
The result is shown in the display in mg/l Ammonia as N.
1
2
3
4
Countdown 1
3:00
start:
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown 2
15:00
start:
6
2
Содержание Lovibond MD 610
Страница 14: ...8 MD 610_2d 11 2019...
Страница 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Страница 87: ...81 MD 610_2d 11 2019...
Страница 95: ...89 MD 610_2d 11 2019...
Страница 279: ...273 MD 610_2d 11 2019 Notes 1 SO3 Na2 SO3 Reagent Formof reagent Quantity Order No SULFITE LR Tablet 100 518020BT...
Страница 306: ...300 MD 610_2d 11 2019...
Страница 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Страница 310: ...304 MD 610_2d 11 2019...
Страница 362: ...356 MD 610_2d 11 2019...
Страница 363: ...357 MD 610_2d 11 2019 Part 3 Enclosure...