334
MD 610_2d 11/2019
After the reaction period is finished proceed as
follows:
9.
Rinsethe vial (24 mm Ø) with the coloured samplefrom
the beaker and fill with 10 ml of the sample.
10. Press
TEST
key.
The batch related method blank is saved.
Press[ ] key,
to go back to mode menu.
Notes:
1. If a new batch of CALCIOtablets is used a calibration of the method blank has to be
performed to optimise the results.
2. Deionisedor tap water
3. If no water free of Calcium is available these ions can be masked by using EDTA.
Preparation: Add 50 mg (a spatula-tipful) EDTAto 100 ml water and dissolve.
4. To achievethe most accurate method blank it is important to adhere exactly to the
sample volume of 100 ml.
Calcium Hardness Method 191 –
Reset method blank to factory calibration
Press[MODE], [Shift] + [4] [0] keys.
Confirm with [ ] key.
The display shows:
Press[Shift] + [2] keys.
The display shows:
<Calibration>
M191 Ca-hardness2T
Reset ?
YES:
Shift + 1
NO:
Shift + 0
<Calibration>
1: M 191 Ca-hardness 2
2: M 191 reset 0 cali.
3: M 170 Fluoride L
prepare Test
press TEST
stored
Mode
0
4
2
Shift
Содержание Lovibond MD 610
Страница 14: ...8 MD 610_2d 11 2019...
Страница 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Страница 87: ...81 MD 610_2d 11 2019...
Страница 95: ...89 MD 610_2d 11 2019...
Страница 279: ...273 MD 610_2d 11 2019 Notes 1 SO3 Na2 SO3 Reagent Formof reagent Quantity Order No SULFITE LR Tablet 100 518020BT...
Страница 306: ...300 MD 610_2d 11 2019...
Страница 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Страница 310: ...304 MD 610_2d 11 2019...
Страница 362: ...356 MD 610_2d 11 2019...
Страница 363: ...357 MD 610_2d 11 2019 Part 3 Enclosure...