94
MD 610_2d 11/2019
COD LR
with Vario Tube Test
3 – 150 mg/l O
2
Insert the adapter for 16 mm Ø vials.
1.
Open one white capped reaction vial and add
2 ml
deionised water
(this is the blank (Note 1)).
2. Open another white capped reaction vial and add
2 ml of the water sample
(this is the sample).
3.
Closethe vialswith the cap tightly. Invert the vial gently
severaltimes to mix the contents.
(CAUTION:The vial will becomehot during mixing!)
4.
Heat the vialsfor
120 minutes
in the preheated reactor
at a temperature of
150°C
.
5.
(CAUTION: The vials are hot!)
Removethe tubesfrom the heating block andallow them
to cool to 60°C or less. Mix the contents by carefully
inverting each tube several times while still warm.
Then allow the tubes to cool to ambient temperature
before measuring. (Note 2).
6.
Place the vial (the blank (Note 3, 4)) in the sample
chamber making sure that the marks are
l
aligned.
7.
Press
ZERO
key.
8.
Removethe vial from the sample chamber.
9.
Place the vial (the sample (Note 3, 4)) in the sample
chamber making sure that the marks are
l
aligned.
10. Press
TEST
key.
The result is shown in the display in mg/l COD.
Zero accepted
prepare Test
press TEST
prepare Zero
press ZERO
Ø 16 mm
3
0
1
Содержание Lovibond MD 610
Страница 14: ...8 MD 610_2d 11 2019...
Страница 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Страница 87: ...81 MD 610_2d 11 2019...
Страница 95: ...89 MD 610_2d 11 2019...
Страница 279: ...273 MD 610_2d 11 2019 Notes 1 SO3 Na2 SO3 Reagent Formof reagent Quantity Order No SULFITE LR Tablet 100 518020BT...
Страница 306: ...300 MD 610_2d 11 2019...
Страница 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Страница 310: ...304 MD 610_2d 11 2019...
Страница 362: ...356 MD 610_2d 11 2019...
Страница 363: ...357 MD 610_2d 11 2019 Part 3 Enclosure...