326
MD 610_2d 11/2019
Data transfer of
results of a selected Code No. range
Press[MODE], [Shift] + [2][2] keys.
Confirm with [ ] key.
The display shows:
Enter numeric code number (up to 6 places) for the first
required Code No., e.g.: [Shift] + [1].
Confirm with [ ] key.
The display shows:
Enter numeric code number (up to 6 places) for the last
required Code No., e.g.: [Shift] + [1][0].
Confirm with [ ] key.
The display shows:
Press [ ] key and all stored results in the selected code
number range are transmitted.
After transfering the photometer goesback to mode menu
automatically.
Note:
1. It is possible to cancel the entry by [ESC].
2. If you want to transfer only results of one code number enter the same code number
twice.
3. If you want to transfer all resultswithout code no. (code no. is 0) enter Zero [0] twice.
from
000001
to
000010
Start:
cancel: ESC
<Data Transfer>
sorted: Code-No.
from _ _ _ _ _ _
to _ _ _ _ _ _
Mode
2
2