113
MD 610_2d 11/2019
Notes:
1. For determination of total Copper digestion is required.
2. Extremely acid water samples (pH 2 or less) must be adjusted between pH 4 and pH 6
before the reagent is added (with 8 mol/l Potassiumhydroxide solution KOH).
Caution: pH valuesabove 6 can lead to Copper precipitation.
3. Accuracy is not affected by undissolved powder.
4. Interferences:
Cyanide, CN
-
Cyanide prevents full colour development.
Add 0.2 ml Formaldehydeto 10 ml water sample and wait for a
reaction time of 4 minutes (Cyanide is masked). After this perform
test as described. Multiply the result by 1.02 to correct the sample
dilution by Formaldehyde.
Silver, Ag
+
If a turbidity remains and turns black, silver interference is likely.
Add 10 drops of saturated Potassiumchloride solution to 75 ml of
water sample. Filtrate through a fine filter. Use10 ml of the filtered
water sample to perform test.
Reagent
Form of reagent/Quantity
Order-No.
VARIO Cu 1 F10
Powder Pack / 100
530300