158
MD 610_2d 11/2019
Iron, total LR
with Liquid reagent
0.03 – 2 mg/l Fe
2+/3+
Digestion procedure
for the determination of total iron.
Total iron consists of soluble, complexed and suspended
iron. Do not filter the sample but ensure the sample is
homogeneous by vigorously shaking immediately prior to
sampling. For Total Soluble (including all complexed) filtra-
tion will be necessary.
This procedure requires equipment and reagents not in-
cluded in the standard test pack supplied.
1.
Filla clean100-ml-Erlenmeyerflask with
50 ml homog-
enized sample.
2.
Add
5 ml 1:1 Hydrochloricacid
and
one spoon KP962
(Ammonium Persulphate Powder)
.
3.
Boil for
20 minutes
, maintaining the sample volume
above 25 ml with deionised water.
4.
Cool the sample to room temperature.
5.
Fill the vial with drops of the same size by holding the
bottle vertically and squeezeslowly:
1 drop KS135 (Phenolphthalein Substitute Indica-
tor)
6.
Add drops of
KS144 (Calcium Hardness Buffer)
, one
drop at a time
with mixing
, until a pink/red colour just
appears.
7.
Fill the sample up to 50ml with deionised water.
8.
Fill a clean vial (24 mm Ø) with
10 ml of the water
sample
, closetightly with the cap.
9.
Placethe vial in the sample chamber making sure that
the marks
are aligned.
2
5
2
Содержание Lovibond MD 610
Страница 14: ...8 MD 610_2d 11 2019...
Страница 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Страница 87: ...81 MD 610_2d 11 2019...
Страница 95: ...89 MD 610_2d 11 2019...
Страница 279: ...273 MD 610_2d 11 2019 Notes 1 SO3 Na2 SO3 Reagent Formof reagent Quantity Order No SULFITE LR Tablet 100 518020BT...
Страница 306: ...300 MD 610_2d 11 2019...
Страница 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Страница 310: ...304 MD 610_2d 11 2019...
Страница 362: ...356 MD 610_2d 11 2019...
Страница 363: ...357 MD 610_2d 11 2019 Part 3 Enclosure...