196
MD 610_2d 11/2019
Nitrogen, total HR
with Vario Tube Test
5 – 150 mg/l N
Insert the adapter for 16 mm Ø vials.
1.
Open two TN Hydroxide HRdigestion vials
and add
the contentsof
one Vario TN Persulfate Rgt. Powder
Pack
(Note 2, 3).
2.
Add
0.5 ml deionised water
to the preparedvial (this
is the blank, Note 4, 5).
3.
Add
0.5 mlof the water sample
to theother preparedvial
(this is the sample).
4. Close the vials with
the caps and shake to mix the
contents (at least 30 seconds, Note 6).
5.
Heat the vialsfor
30 minutes
in the preheated reactor
at a temperature of
100°C
(Note 7).
6.
After 30 Minutes remove the vials from the reactor.
(CAUTION: The vials are hot!)
Allow the vials to cool to room temperature.
7.
Open the cooled digestion vialsand add the contentsof
one
Vario TN Reagent A Powder Pack
to each vial (Note 2).
8. Close the vials with
the caps and shake to mix the
contents (at least 15 seconds).
9. Press
[
]
key.
Wait for a
reaction period of 3 minutes.
After the reaction period is finished proceed as follows:
10. Open the digestion vials and add the contents of
one
Vario TN Reagent B Powder Pack
to each vial (Note 2).
Countdown
3:00
start:
1
2
3
4
Ø 16 mm
8
1
2
Содержание Lovibond MD 610
Страница 14: ...8 MD 610_2d 11 2019...
Страница 15: ...9 MD 610_2d 11 2019 Part 1 Methods...
Страница 87: ...81 MD 610_2d 11 2019...
Страница 95: ...89 MD 610_2d 11 2019...
Страница 279: ...273 MD 610_2d 11 2019 Notes 1 SO3 Na2 SO3 Reagent Formof reagent Quantity Order No SULFITE LR Tablet 100 518020BT...
Страница 306: ...300 MD 610_2d 11 2019...
Страница 307: ...301 MD 610_2d 11 2019 Part 2 Instrument Manual...
Страница 310: ...304 MD 610_2d 11 2019...
Страница 362: ...356 MD 610_2d 11 2019...
Страница 363: ...357 MD 610_2d 11 2019 Part 3 Enclosure...