175
MD 610_2d 11/2019
12. Press
TEST
key.
Wait for a
reaction period of 3 minutes.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Manganese.
Notes:
1. The following substancesinterfer with this test:
Calcium
> 500mg/l
Sodium
> 500mg/l
Nickel
> 0.5 mg/l
Iron
> 5 mg/l
Chromium > 5 mg/l
Reagent
Form of reagent/Quantity
Order-No.
KS265 – Manganese Reagent A
Liquid reagent / 30 ml
56L026530
KS266 – Manganese Reagent B
Liquid reagent / 30 ml
56L026630
KS304 – Manganese Reagent C
Liquid reagent / 30 ml
56L030430
Zero accepted
prepare Test
press TEST
Countdown
3:00