Corners
A
3.3 nM
3.3
nM
3.3 nM
BUF
BUF
3.3
nM
3.3
nM
3.3
nM
B
BUF
BUF
BUF
BUF
BUF
BUF
BUF
BUF
C
150
µL:
1 nM
0.5
nM
0.25
nM
0.125
nM
0.0625
nM
BUF
Sensitivity
/Linearity
D
1 nM
0.5
nM
0.25
nM
0.125
nM
0.0625
nM
200
µL:
1 nM
BUF
E
1 nM
0.5
nM
0.25
nM
0.125
nM
0.0625
nM
BUF
F
1 nM
0.5
nM
0.25
nM
0.125
nM
0.0625
nM
G
BUF
BUF
BUF
BUF
BUF
BUF
BUF
BUF
Corners
H
3.3 nM
3.3
nM
3.3 nM
BUF
BUF
3.3
nM
3.3
nM
3.3
nM
Fluorescence Polarization (FP) Test:
The plate used for testing Corners/Sensitivity/Linearity of the Top optics can
also be used for this test.
l
Pipette 200 μL of the (green) polarization buffer (BUF) into wells A6–H6.
l
Pipette 200 μL of the green high polarization reference (HPR) into wells A7–B7.
l
Pipette 200 μL of the green low polarization reference (LPR) into wells A8–H8.
1
2
3
4
5
6
7
8
9
10
11
12
A
BUF
HPR
LPR
B
BUF
HPR
LPR
C
BUF
LPR
D
BUF
LPR
E
BUF
LPR
F
BUF
LPR
G
BUF
LPR
H
BUF
LPR
128 | Chapter 8: Instrument Qualification Procedures
BioTek Instruments, Inc.
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...