Replace the Syringe
Refer to
on page
After using Gen5 to move the syringe into its maintenance position:
1. Using your fingers, unscrew the bottom thumbscrew that secures the syringe,
underneath the bracket. Retain this bottom thumbscrew; it is needed for the
replacement syringe.
2. Unscrew the top thumbscrew to disengage the syringe from the valve.
3. Remove the new syringe from its protective box.
4. Hold the syringe vertically with the threaded end at the top. Screw the top of the
syringe into the bottom of the syringe valve. Finger-tighten only.
5. Carefully pull down the bottom of the syringe until it rests inside the hole in the
bracket.
6. Pass the thumbscrew (used to hold the old syringe) up through this hole and thread it
into the bottom of the syringe. Hold the syringe from rotating while tightening the
thumbscrew. Finger-tighten only.
7. In Gen5, select
System > Instrument Control > Synergy H1
.
8. Click the
Prime
tab and click
Initialize
.
72 | Chapter 6: As-Needed Maintenance
BioTek Instruments, Inc.
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...