When each dispense step is finished, you will weigh the plate, record the
weight, tare the balance with the plate on it, and then place the plate back
on the carrier for the next step.
7. Initiate a plate read. Gen5 will prompt you to empty the tip priming trough.
8. When ready, proceed with the experiment. The sequence is as follows:
l
80 μL/well is dispensed to columns 1–4.
l
When prompted, remove the plate and weigh it. Record the weight and tare the
balance. Place the plate on the carrier.
l
20 μL/well is dispensed to columns 5–8.
l
When prompted, remove the plate and weigh it. Record the weight and tare the
balance. Place the plate on the carrier.
l
5 μL/well is dispensed to columns 9–12.
l
When prompted, remove the plate and weigh it. Record the weight.
l
Manually pipette
150 μL
of deionized or distilled water into all 12 columns, on top
of the green test dye solution.
l
Place the plate on the carrier for the shake and read steps.
9. When the experiment is complete, save the file with an identifying name.
10. Remove the plate from the carrier and set it aside.
11. Repeat the procedure using the
Synergy H1 Disp 2 Test
protocol and a new
microplate.
12. When the tests are complete:
l
Prime both dispensers with at least 5000 µL of deionized water to flush out the dye
solution.
l
Refer to the instructions on page
to perform calculations and determine
pass/fail.
l
Test descriptions are provided on page
.
136 | Chapter 8: Instrument Qualification Procedures
BioTek Instruments, Inc.
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...