Gen5 System Test Report
Reader: Synergy H1 (Serial Number: 1506126)
Basecode: P/N 8040200 (v2.00)
Gen5 Version: 2.07.17
Date and Time: 6/12/2015 2:28:12 PM
User: 4205
Company: BioTek
Comments:
Test Results
SYSTEM TEST PASS
Operator ID:______________________________________________________________
Notes:____________________________________________________________________
__________________________________________________________________________
SYSTEM SELF TEST
8040200 Version 2.00 1506126 S1 S2 S3
1110 0100
DMF
Voltage Reference Test Min Low High Max
Mono System Flash 1412 1750 2318 3098
Filter System Flash 2338 2922 3507
Switched 24V Power 1886
ABSORBANCE
Optics Test Ref Meas Gain Resets
#1:230 1.91 2
Light 13389 39922
Dark 10613 10655
Delta 2776 29267
#2:285 2.91 8
Light 13064 39639
Dark 10647 10679
Delta 2417 28960
#3:405 1.62 8
Light 12886 39571
Dark 10621 10641
Delta 2265 28930
#4:630 1.65 8
Light 12611 39827
Dark 10622 10642
Delta 1989 29185
#5:800 2.12 4
Light 12502 39891
Dark 10626 10657
Delta 1876 29234
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...