Kinetic Assays Using the Continuous Shake Feature
This recommendation applies only to Synergy H1 basecode software versions
lower than 2.00.
A Gen5 experiment that specifies the following parameters may not run successfully on the
Synergy H1: Continuous Shake, Kinetic Interval greater than 15 minutes
When the experiment is initiated, plate shaking will begin, but shaking may stop
prematurely with no error message.
One suggested workaround is to shorten the kinetic interval. For example, if your desired
experiment is 25 kinetic reads with 60-minute intervals, use 100 kinetic reads with 15-
minute intervals.
Another suggestion is to perform multiple Shake steps and then a Read step with the
Discontinuous Kinetic Procedure feature enabled. For example, if your desired experiment
looks like this:
do this (example assumes the Read step takes one minute):
Synergy H1 Operator's Manual
Recommendations for Optimum Performance | 41
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...