Synergy H1 FI_T_MUB.prt
Parameter
Setting
Plate Type
"Costar 96 black opaque" (#3915)
Read Step 1
Kinetic loop
Run time 0:01:00, Interval 0:00:04 (16 reads)
Detection Method
Fluorescence intensity
Read Type
Endpoint / Kinetic
Optics Type
Filters
Step Label
"Sensitivity Read"
Read well
D7
Filter Set
1 ("Blue")
Excitation
360/40 nm
Emission
460/40 nm
Optics Position
Top 400 nm
Gain
Auto, Scale to High Wells, D7, 80000
Read Speed
Normal
Delay after plate movement
350 msec
Measurements per data point
40
Lamp Energy
Low (faster)
Read Height
7.00 mm
Read Step 2
Kinetic loop
Run time 0:01:30, Interval 0:00:06 (16 reads)
Detection Method
Fluorescence intensity
Read Type
Endpoint / Kinetic
Optics Type
Filters
Step Label
"Sensitivity Read Buffer"
Read well
C9, D9, E9
Filter Set
1 ("Blue")
Excitation
360/40 nm
Emission
460/40 nm
Optics Position
Top 400 nm
Gain
Auto, Use first filter set gain from FIRST Read
Step
Read Speed
Normal
Delay after plate movement
350 msec
Synergy H1 Operator's Manual
Fluorescence Testing | 95
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...