Parameter
Setting
Measurements per data point
100
Lamp Energy
Low (faster)
Read Height
7.00 mm (Top only)
Read Step 3
Detection Method
Fluorescence intensity
Read Type
Endpoint / Kinetic
Optics Type
Filters
Step Label
"Corners Read"
Read well
Full Plate (to support Gen5 v1.11)
Filter Set
1 ("Green")
Excitation
485 nm
Emission
528 nm
Optics Position
Top (or Bottom)
Gain
Auto, Scale to High Wells, A3, 50000
Read Speed
Normal
Delay after plate movement
350 msec
Measurements per data point
100
Lamp Energy
Low (faster)
Read Height
7.00 mm (Top only)
Read Step 4
Detection Method
Fluorescence intensity
Read Type
Endpoint / Kinetic
Optics Type
Filters
Step Label
"Linearity Read"
Read well
C1-F5
Filter Set
1 ("Green")
Excitation
485 nm
Emission
528 nm
Optics Position
Top (or Bottom)
Gain
Auto, Scale to High Wells, C1, 50000
Read Speed
Normal
Delay after plate movement
350 msec
Measurements per data point
100
Lamp Energy
Low (faster)
92 | Chapter 7: Instrument Qualification Process
BioTek Instruments, Inc.
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...